Precision Premium Oven 2-1
Thermo Scientific
Section 2
Specifications
Performance Characteristics
Operating Range . . . . . . . . . . . . . . . . . . .50 to 275°
Average Uniformity @ 200°*
Forced Air Models . . . . . . . . . . . . . . . . . . . . .±3°C
Gravity Models . . . . . . . . . . . . . . . . . . . . . . .±4°C
Control Resolution . . . . . . . . . . . . . . . . . . . . . . .1°C
Control Sensitivity . . . . . . . . . . . . . . . . . . . . .±0.5°C
Recovery Time @ 200°C**
PR305045M, -40M & -40MCN . . . . . .1.0 minutes
PR305045G, -40G & -40GCN . . . . . . .2.0 minutes
PR305055M, -50M & -50MCN . . . . . .2.0 minutes
PR305055G, -50G & -50GCN . . . . . . .3.0 minutes
PR305065M, -60M & -50MCN . . . . . .2.5 minutes
PR305065G, -60G & -60GCN . . . . . . .4.0 minutes
Rise Time to 275°C
PR305045M, -40M & -40MCN . . . . . . .70 minutes
PR305045G, -40G & -40GCN . . . . . . .40 minutes
PR305055M, -50M & -50MCN . . . . . .80 minutes
PR305055G, -50G & -50GCN . . . . . . .80 minutes
PR305065M, -60M & -50MCN . . . . . .80 minutes
PR305065G, -60G & -60GCN . . . . . .100 minutes
Air Exchanges per Hour*
PR305045M, -40M & -40MCN . . . . . . . . . . . . .43
PR305045G, -40G & -40GCN . . . . . . . . . . . . . .24
PR305055M, -50M & -50MCN . . . . . . . . . . . . .29
PR305055G, -50G & -50GCN . . . . . . . . . . . . . .16
PR305065M, -60M & -50MCN . . . . . . . . . . . . .22
PR305065G, -60G & -60GCN . . . . . . . . . . . . . . .2