To access the electronics compartment, proceed as follows (refer to Figure
6-1):
1. Disconnect power cord from the electrical outlet.
2. Open the chamber door. Carefully remove and retain hardware from
door hinges (case side). Set door aside.
3. Slide the oven forward until the front of the bezel (control panel) is at
least three inches from the edge of the bench top (or the oven feet are
at the edge of the bench).
4. Prop up the oven front by placing a shim under each front foot. Use
shims between 1½ and 2 inches in thickness.
5. Remove the screws securing bezel from bottom of oven.
6. Slide the oven back on the table plus a few inches (to set the bezel on
the bench) and rotate the bottom of the bezel out from the oven. The
top clips will come loose but the wiring will still be connected.
7. Carefully set the bezel on the bench.
6-4
Precision Premium Oven
Thermo Scientific
Section 6
Service
Accessing Electronics
Compartment
Heater
Electronics
Compartment
(Open)
Bezel/Control
Panel
Heater
Cover
Front
Face
Fascia
Bezel
Bottom
Screws
Figure 6-1.