background image

      This is the delay time setting for the low alarm. A difference

between the high alarm and low alarm delay that should be noted
at this point is that when you select a value for the high alarm
delay and then turn power off, that value will appear again when
power is turned on again. The low alarm delay however will always
be 20 minutes when power is turned on. The reason for this, if the
delay is set for too short of a time and the compressor goes off, it
will be approximately 10 minutes before the compressor turns back
on. This is due to an internal time delay within the compressor.
The user can change this 20 minutes to a value desired, but when
power is turned off and on again, it will reset to 20 minutes again.
Use the up or down arrow keys to select your time value and press
ENTER. 

      The displays will change to the following:

      This is the value for the alarm band, the amount of temperature

deviation allowed before either the high or low delay timers are
started. Use the up or down arrow keys to select your value and
then press ENTER and the displays will return to their normal
operation.

8.   CALENDAR - The clock key allows the hours, minutes, and day of

the internal clock to be set. To set the internal clock, select
Configuration mode 2. To set the date, month, and year, follow these
instructions:

A.   Press and hold the up or down arrow keys and then press the

ENTER key and the displays will change to the following:

      Use the key to select 02 and then press ENTER. The displays will

change to the following:

      A number other than 01 might be shown.

      This is the day of the month selection. Use the up or down arrow

keys to select the correct date and then press ENTER and the
displays will change to the following:

Plant Growth Chamber

3-9

Thermo Scientific

Section 3

Operation and Calibration

Summary of Contents for PR505750L

Page 1: ...to register your warranty www thermoscientific com labwarranty User Manual Thermo Scientific Precision Model 818 Incubator Plant Growth Chamber Microprocessor Controlled Operating and Maintenance Manu...

Page 2: ...g 0 40096 4 3 15 Original new cabinet size technical data ccs Rev ECR ECN Date Description By Preface Plant Growth Chamber i Catalog Number Model Volume L cu ft Voltage PR505755L 3759 504 17 8 120V PR...

Page 3: ...ided by the equipment design may be impaired s Warning Hot warm surface s can occur Depending on ambient termperature and settings surface can become warm up to 60 C s Material in this manual is for i...

Page 4: ...oduct is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with the WEEE symbol Thermo Fisher Scientific has contracted with one...

Page 5: ...eds for spare or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Thermo Scientific...

Page 6: ...llable service at your location Calibration after a warranty repair is covered under the warranty Damage resulting from use of improper quality water chemicals or cleaning agents detrimental to equipm...

Page 7: ...8 Installation 2 1 Safety Precautions 2 2 Operation and Calibration 3 1 To Lock the Keyboard 3 6 To Unlock the Keyboard 3 7 Adjusting the Intake and Exhaust Ports 3 10 Service 4 1 Troubleshooting 4 1...

Page 8: ...1 Visible Loss or Damage Note any external evidence of loss or damage on the freight bill or express receipt and have it signed by the carrier s agent Failure to adequately describe such external evid...

Page 9: ...viation from set value spatial 27 point measurement according to DIN 12880 Temperature stability Maximum temperature difference at chamber center point 1 point measurement Energy consumption Heat up t...

Page 10: ...s even air distribution throughout the chamber and ensures temperature uniformity The program mode of PR505750L PR505755L plant growth chamber allows the user to program into memory 2 sets of values f...

Page 11: ...0 ON SUNDAY 1 AM 00 20 0 OFF 3 PM 00 40 0 ON FRIDAY 2 AM 00 10 0 ON 4 PM 00 20 0 OFF SAMPLE SCHEDULE SHEET HOURS MINUTES SETPOINT LIGHT HOURS MINUTES SETPOINT LIGHT AM PM TEMPERATURE ON OFF AM PM TEM...

Page 12: ...quence 5 Temperature Display This four digit display indicates the growth chamber chamber temperature Also during programming it displays the value prompted by the time display 6 Heater On Lamp This l...

Page 13: ...a setpoint temperature if the continuous mode of operation has been selected 13 Up Down Arrow keys These keys are used to increase or decrease a desired value 14 REPEAT KEY This key is a time saver ke...

Page 14: ...w temperature alarm condition This alarm relay will operate just as the audible alarm would The contact itself is an isolated form C normally open normally closed dry contact This contact is to be use...

Page 15: ...ommended main use of this is with a chart recorder having an input impedance of at least 1M ohm The scaled temperature output change is 10mV C When the growth chamber is operating at a negative temper...

Page 16: ...F as the compressor thermo overload will be tripped and will result in a wide band control cycle of approximately 4 C Such cycling should not be misinterpreted as a malfunction of the electronic cont...

Page 17: ...ker in use Caution Be sure the power supply is the same voltage as specified on the nameplate s When loading the growth chamber a space of 1 2 must be allowed between adjacent items in the chamber Thi...

Page 18: ...thin 15 seconds the displays will return to their previous state Using either the up or down arrow keys the user can select the correct hour of the day After the correct hour has been selected ENTER m...

Page 19: ...ect the day While using the up or down arrow keys observe the seven day of the week lamps for selecting the correct day After the correct day has been selected press ENTER and the growth chamber will...

Page 20: ...ke some time to study the sample schedule sheet provided on page 1 4 For each day there are eight items which must be programmed and are as follows Hour 1 AM or PM Minute 1 Setpoint Temperature 1 Ligh...

Page 21: ...or setting the clock but note the AM PM lamps After your selection is displayed press the ENTER key and the displays will change to the following Use the up or down arrow keys to select your choice fo...

Page 22: ...e for the setpoint you want the growth chamber to switch to a your second hour and minute you have chosen After your selection is displayed press the ENTER key and the displays will change to the foll...

Page 23: ...out a half second the displays will return to normal operation E All of the days program schedules will be the same as the present day 5 Lock Key The keyboard may be locked to prevent inadvertent chan...

Page 24: ...e CALIBRATE key should be used ONLY to match a stable growth chamber s actual temperature to the calibrated thermometer Place a calibrated thermometer or the probe of one at the center of the third sh...

Page 25: ...r 2 minutes the low alarm temperature will go off along with the compressor This is not a real alarm failure If the door is left open long enough for this to occur then the door must be closed sooner...

Page 26: ...displays will change to the following This is the value for the alarm band the amount of temperature deviation allowed before either the high or low delay timers are started Use the up or down arrow...

Page 27: ...amber will return to normal operation at this point The exhaust port is the top port on the door Open it slightly by turning it counterclockwise to permit exhausting of chamber air The intake port is...

Page 28: ...ubators Your equipment has been thoroughly tested and calibrated before shipment Regular preventive maintenance is important to keep your VOJU functioning properly The operator should perform routine...

Page 29: ...ic temperature control First the symptoms and possible causes will be listed for quick reference Possible solutions will be suggested later Warning Hazardous high voltage conditions exist inside the c...

Page 30: ...1 Low line voltage less than 110V see Solution 9 RTD Temperature Probe see Solution 3 Step D and or Solution 4 Power Supply PCB Assembly see Solution 3 Overcharged compressor see Solution 11 Differen...

Page 31: ...ndicates an open circuit it may or may not be open In series with the heater is a thermal overload switch It might be that this thermal overload is open and not the heater To check the heater and ther...

Page 32: ...urements will be DC voltage measurement so a DC scale of at least 10 volts should be selected Connect the negative lead to TP2 This is DC ground Also locate the terminal block TB1 B This first section...

Page 33: ...lamp on the front panel It should be on continuously not flashing 2 Measure the voltage at U1 Pin 2 with respect to TP2 It should be no greater than 4 volts DC 3 Measure the DC voltage across R4 sinc...

Page 34: ...racket attached on the rear wall and near the top of the chamber Two tube shaped clips securely support the RTD probe into place The RTD must be inserted fully into its support bracket 6 Constantly Ex...

Page 35: ...chamber temperature will be affected Correct as necessary Caution The growth chamber should not be operated in an environment where the ambient temperature exceeds 90 F as the compressor thermo overlo...

Page 36: ...ambient temperatures or in any manner that increases moisture buildup within the chamber an appropriate defrosting schedule must be determined Defrosting is accomplished by turning the growth chamber...

Page 37: ...n the RS232 PCB assembly see sketch 6 Replace the control panel and secure with the eight 8 screws The RS232 PCB provides a bi directional communication port that will allow the user to monitor the pe...

Page 38: ...ication software selected to be used B Status Update Time Interval This is the time between information being sent from the growth chamber until it is received by the computer C Status Display Type Se...

Page 39: ...shown in the display Choose the number appropriate for the Baud Rate desired 3 300 6 600 12 1200 24 2400 48 4800 96 9600 Press the UP or DOWN key until the display shows your selection Press ENTER The...

Page 40: ...into a spreadsheet 1 Used for a multi line format with English headings with continuous screen 2 Used for the format as 1 but preceded by a clear screen command used by terminals or computers running...

Page 41: ...1 COM Port Selected by user Format ASCII The growth chamber transmits information on its RS232 port as long as the status update time interval is not 0 0 If you start your communication software with...

Page 42: ...d ENTER Return Example 1 To change Setpoint Temperature to 25 0 C type T25 0 CR Example 2 To change Setpoint Temperature to 8 C type T 8 CR A Actual Temperature The actual temperature within the growt...

Page 43: ...5 5 0 Minutes 6 10 0 Minutes 7 30 0 Minutes 8 60 0 Minutes V EPROM Version Number The growth chamber software version level D Status Display Type Selection of format that data is sent to your compute...

Page 44: ...requency 120V 60Hz 230V 60Hz 230 50Hz Power rating 720 W 720 W 720 W Max current 6 8 A 3 4 A 3 4 A Plug type 120V Nema 5 15P 230V Nema 6 15P China Plug Max BTU output 2220 650W Access port diameter mm...

Page 45: ...gory used for instruments in installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial laboratories the expected transient overvolt...

Page 46: ...TO CPU 3176761 CABLE ASS Y CPU TO INT 3176762 CABLE ASS Y INT TO PWR 3176763 BALLAST ASS Y 3167073 FUSE KIT 10 AMP 115V 3167218 FUSE HOLDER KIT 115V 3167216 FUSE HOLDER KIT 230V 3167217 CORD W PLUG 1...

Page 47: ...e Parts Accessories Name Part Number Glass Thermometers 3173710 RS232 Interface Kit 3166245 Digital Chart Recorder 3166207 Chart Paper 100 sheets 3174255 Access Port Stopper 3167215 Shelf includes fal...

Page 48: ...Plant Growth Chamber 9 1 Thermo Scientific WIRING DIAGRAM 818 INCUBATOR 115V 3167073 3177654 3166245 3176817 3176818 3176825 3176756 3176816 3175264...

Page 49: ...9 2 Plant Growth Chamber Thermo Scientific 5 1 5 0 1 8 725 9...

Page 50: ...SCHEDULE SHEET HOURS MINUTES SETPOINT LIGHT HOURS MINUTES SETPOINT LIGHT AM PM TEMPERATURE ON OFF AM PM TEMPERATURE ON OFF WEDNESDAY SATURDAY FRIDAY THURSDAY TUESDAY MONDAY SUNDAY...

Page 51: ...mo Fisher Scientific and its subsidiaries Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for deta...

Reviews: