PROGRAM mode of the 3758/3759 Plant Growth Chamber allows
the user to program in two sets of time, setpoint temperature, and light
conditions for each day of the week. This is started by selecting
PROGRAM mode pressing the SET key, then pressing the
PROGRAM key. The displays and the indicator lamps will change to
the following:
Use the up or down arrow keys to select your choice for the first hour
for Sunday.
There is no separate selection for AM/PM here as there is for setting
the clock, but note the AM/PM lamps.
After your selection is displayed, press the ENTER key and the displays
will change to the following:
Use the up or down arrow keys to select your choice for the first
minutes for Sunday. After your selection is displayed, press the
ENTER key and the displays will change to the following:
Use the up or down arrow keys to select your choice for the setpoint
you want the growth chamber to switch to at your first hour and
minute you have chosen. After your selection is displayed, press the
ENTER key and the displays will change to the following:
3-4
Plant Growth Chamber
Thermo Scientific
Section 3
Operation and Calibration