Routine Operation
Running a protocol
Thermo Fisher Scientific
Thermo Scientific PikoReal Real-Time PCR System User Manual
31
Note
Protocol files on the USB memory stick can have a name that is
up to eight characters long.
▲
Note
Thermo Fisher Scientific does not guarantee that all USB memory
devices are compatible with the instrument.
▲
5.
The protocol can be viewed in list mode or graph mode, and
browsed using the Left/Right or the Up/Down arrow keys,
correspondingly.
6.
After selecting the view mode (list or graph), start a run by selecting
START
and pressing Enter.
With a 24-well PikoReal Real-Time PCR System, the users will be
presented with consumable and volume options. With a 96-well
PikoReal Real-Time PCR System, the users will only be presented with
volume options.
7.
Select the consumable (that is, the vessel type) for a 24-well
PikoReal Real-Time PCR System. Select either:
●
< 16 TUBES
if using 8–16 individual 0.2 ml tubes
or
●
> 16 TUBES OR PLATE
if using a 24-well Piko PCR Plate.
Caution
To avoid risk of contaminating the instrument, always use
high-quality plastic consumables, such as UTW consumables. For more
information, refer to Chapter 8: “
Ordering Information
”.
▲