background image

263663-001   Rev. B   07-09

9 0 0 1 : 2 0 0 8

Water Analysis Instruments

North America 

166 Cummings Center 

Beverly, MA 01915  USA 

Toll Free: 1-800-225-1480 

Tel: 1-978-232-6000 

Dom. Fax: 1-978-232-6015 

Int’l Fax: 978-232-6031

Europe

P.O. Box 254, 3860 AG Nijkerk  

Wallerstraat 125K, 3862 CN Nijkerk,  

Netherlands 

Tel: (31) 033-2463887  

Fax: (31) 033-2460832

Asia Pacific

Blk 55, Ayer Rajah Crescent 

#04-16/24, Singapore 139949 

Tel: 65-6778-6876 

Fax: 65-6773-0836

www.thermo.com/water

© 2009 Thermo Fisher Scientific Inc.  All rights reserved.

Summary of Contents for Orion COD165

Page 1: ...Thermo Scientific Orion COD165 Thermoreactor User Guide ...

Page 2: ...nk ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Star Navigator 21 Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are trademarks of Thermo Fisher Guaranteed Success and The Technical Edge are service marks of Thermo Fisher PerpHecT meters are protected by U S patent 6 168 707 PerpHecT ROSS electrodes are protected by U S patent 6 168...

Page 3: ...Connecting the Power Supply 3 Start Up 3 Operation Controls 3 Power Interruptions 4 Auto Shutoff Feature 4 Safety Features 4 Operation 5 Running a Cycle 5 Maintenance 6 Replacing a Fuse 6 Disposing of the Instrument 6 Verification of the Heating Block Temperature 6 Customer Services 7 Specifications 7 Electrical Schematic 8 Notice of Compliance 9 Assistance 9 Warranty 9 Ordering Information 10 Cha...

Page 4: ...II Thermo Scientific Orion COD165 Thermoreactor User Guide ...

Page 5: ...ass containers that can withstand a temperature greater than 165 C should be used with the COD165 thermoreactor Always unplug the COD165 thermoreactor before cleaning it The heating plate must be cool before the instrument is cleaned Use a damp cloth with nonflammable non abrasive detergents The COD165 thermoreactor operates at high temperatures and is designed for use with glass vials containing ...

Page 6: ...s the COD vials allows for very good thermal homogeneity at the selected temperatures The temperature of the heating block is microprocessor controlled The Pt100 temperature probe does not require any calibration because when the instrument is turned on a self test is performed The COD165 thermoreactor is equipped with a voltage selector for either 115 V or 230 V It is accessible on the rear panel...

Page 7: ...e temperature press the Start Temp key and select desired temperature The temperature setting will scroll down through the four temperatures and the corresponding LED will light Note It is possible to select different temperatures if the programmed temperature has not been reached and the timer is not started Once timer has started this key is not operative If the selected temperature is lower tha...

Page 8: ...t will sound continuously for 5 seconds at the same time the heating element and the corresponding time and temperature LED will turn off The instrument is now ready for a new work cycle The instrument will register a black out or a lack of power supply as a manual stop of the cycle The work cycle will have to be restarted The COD165 thermoreactor will shut down once the timer cycle is complete If...

Page 9: ...me if the selected temperature has not been reached yet and the timer has not started Once the selected temperature has been reached a beep will sound intermittently for five seconds After the beep the LED corresponding to the selected time will start to blink indicating that the timer has started The end of the cycle will be indicated when the beeper sounds continuously for five seconds At the co...

Page 10: ...be opened with a screwdriver See the Electrical Schematic section for the fuse type Disposal of this unit should comply with all local state and federal regulations The electronic temperature controls within the COD165 thermoreactor are designed to provide excellent stability and eliminate temperature overruns or oscillations The temperature sensor does not require calibration as a self calibratio...

Page 11: ...ator PID microprocessor Precision 0 2 end of scale Temperature Probe Pt100 class A Temperature Probe Calibration Automatic by software Heating Block Temperature Stability 0 5 C Heating Block Temperature Homogeneity 0 5 C Heating Block Temperature Precision 1 C Over temperature Protection On the block Performance from 20 C to 150 C 12 minutes Reaching of Set Temperature Signal Acoustic During Time ...

Page 12: ...ific Orion COD165 Thermoreactor User Guide 1 5 x 20 mm slow blow fuse 5 A 2 General switch 3 Voltage changer electronic card 4 Safety thermostat 5 Electrical resistance 6 Pt100 temperature probe Electrical Schematic ...

Page 13: ...rference at his own expense The digital apparatus does not exceed Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications Le present appareil numerque n emetpas de bruits radioelectriques depassant les limites applicables aux appareils numeriques de la class A prescites dans le Reglement sur le broui...

Page 14: ...L bottle CODL00 COD test kit 0 to 150 ppm 25 tests EPA approved CODH00 COD test kit 0 to 1500 ppm 25 tests EPA approved CODHP0 COD test kit 0 to 15000 ppm 25 tests AC2V16 16 mm vials 10 pack AQ4000 AQUAfast IV advanced colorimeter AQ2040 AQUAfast II COD colorimeter AQ3700 AQUAfast III colorimeter with nitrogen phosphate and COD digestion capabilities CODTMP Pt100 probe for temperature measurements...

Page 15: ... 978 232 6000 Dom Fax 1 978 232 6015 Int l Fax 978 232 6031 Europe P O Box 254 3860 AG Nijkerk Wallerstraat 125K 3862 CN Nijkerk Netherlands Tel 31 033 2463887 Fax 31 033 2460832 Asia Pacific Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949 Tel 65 6778 6876 Fax 65 6773 0836 www thermo com water 2009 Thermo Fisher Scientific Inc All rights reserved ...

Reviews: