Thermo Scientific Orion COD165 Thermoreactor User Guide
All the maintenance operations should be performed only after the
instrument has been disconnected from the power supply .
The COD165 thermoreactor is designed to be maintenance-free and only
occasional cleaning needs to be performed with a damp cloth and non-
flammable, non-abrasive detergent .
The heating block is equipped with two fuses connected to the socket,
located on the rear panel . If a fuse replacement is required, the power
supply should be removed and then the fuse panel can be opened with a
screwdriver . See the Electrical Schematic section for the fuse type .
Disposal of this unit should comply with all local, state, and federal
regulations .
The electronic temperature controls within the COD165 thermoreactor are
designed to provide excellent stability and eliminate temperature overruns
or oscillations . The temperature sensor does not require calibration, as a
self-calibration is performed upon power up of the instrument .
The temperature may be monitored using a standard NIST traceable
thermometer with a range of 0 to 250 °C . Place the thermometer in a hole
within the heating block and monitor the temperature .
Maintenance
Replacing the Fuses
Disposing of the
nstrument
Verification of the Heating
Block Temperature