Thermo Scientific Orion COD165 Thermoreactor User Guide
When the unit is switched on, press the Start/Temp key to start the work
cycle . The default option is set at 160 °C and 30 minutes, as shown by the
corresponding LED . To select a different temperature, press the Start/Temp
key until the corresponding LED is lit . To select a different time, press the
Time
key until the corresponding LED is lit .
Note:
It is possible to modify both temperature and time until the
temperature of the heating block reaches the selected temperature and the
run time starts .
When the selected temperature is reached, a beeper will sound
intermittently for 5 seconds . At this time the LED corresponding to the
selected time will start to blink, signalling that the countdown has started .
The end of the countdown will be signaled by a beeper that will sound
continuously for 5 seconds; at the same time the heating element and the
corresponding time and temperature LED will turn off . The instrument is
now ready for a new work cycle .
The instrument will register a black out or a lack of power supply as a
manual stop of the cycle . The work cycle will have to be restarted .
The COD165 thermoreactor will shut down once the timer cycle is
complete . If the power to the instrument is interrupted, the instrument will
automatically end the program that is running and a manual restart
is required .
The COD165 thermoreactor is equipped with an internal thermostat
that will not allow the temperature of the block to exceed 180 °C . If this
temperature is reached the unit will automatically shutdown . The COD165
thermoreactor continuously monitors the block of temperature and if
inconsistent or anomalous readings are observed, the instrument will go
into an alarm mode in which all eight LEDs blink and a beeper sounds .
Auto Shutoff Feature
Safety Features
Power nterruptions