7070iX TRO Analyzer
General Information | 5
Sample Requirements
Refer to
Specifications
for details.
System Description
The Orion 7070iX Total Residual Oxidant Analyzer can be
used on either the discharge side or the generation side of a
Water Treatment System for monitoring the TRO
concentration in the sample stream. The location of the
analyzer is a critical factor in ensuring fast response time and
care should be taken to install the analyzer as close as
possible to the source of the sample.
Analyzer Connections
The water connections to the analyzer include the following:
•
Sample water input – Flow from source to analyzer.
•
Fast-loop sample output – Sample water return.
•
Measured water output – Measurement cell discharge
composed of combined sample and reagent.
Additional connections are for the following:
•
AC power.
•
Remote operation (Electrically Isolated 2 wire Digital
Signal input. + input signal range is + 3 VDC to
+ 5.5 VDC, - side signal is 0 V).
•
Analog data communications (two 4-20 mA).
Sample Inlet
Connection:
The instrument is supplied with a
1/4 - inch OD plastic tube with
quick connect fitting for sample
water inlet
Flow Rate
minimum of 40 mL / minute
Sample Pressure:
5 to 50 psi
Temperature:
41 to 113 °F / 5 to 45 °C
Alkalinity:
Up to 500 ppm as CaCO
3
Max Turbidity:
500 NTUs
Max TSS:
1000 PPM
Max particle size:
130 microns