7070iX TRO Analyzer
Specifications | 33
Specifications
Table 6. Product Specifications
Measurement
Performance
Measuring Range
0.001-15 ppm
Accuracy
Less than 5% of reading or 10 ppb, which ever is greater from
0.001-15 ppm
Resolution
0.001 ppm in all ranges
Response Time
95% within three minutes
Precision
±2% of reading or ±10 ppb, whichever is greater from
0.001-15 ppm
Limit of Detection
0.001 ppm
Method
Based on electrode 9770 ISE EPA approved standard method
Environmental
Ambient Operating
Temperature
5 to 45 ºC (41 to 113 ºF)
Maximum Humidity
85% at 40 °C (104 °F)
Sample
Requirements
Sample Flow
40 mL/minute
Sample Pressure
5-50 psig
Sample Supply
Continuous
Sample Temperature Range
5 to 45 °C (41 to 113 °F)
Max TSS
1000 ppm
Max particle size
130 microns
Max Turbidity
500 NTUs
Chlorine
up to 20 ppm
Sample Inlet/Outlet
Connections
1/4" LLDPE Tubing - length 2 ft min
Drain Tubing
1/8" LLDPE Tubing - length 2 ft min
Sample Streams
One
Construction
Enclosure Integrity
IP54
Enclosure Dimensions
27.9 inches x 18.8 inches x 12.4 inches
709 mm x 478 mm x 315 mm
Shipping Weight
20 kg (45 lbs) without reagent
Electrical
Power Requirements
100 - 240 VAC, 50/60 Hz, 40 watts