Troubleshooting
VI-15
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Displayed
Event Code
Description
Cause
Troubleshooting
System Event/Error Codes - Yellow LED
E004
DEFAULT VALUES
Analyzer has been reset or is new out
of box
Calibrate the analyzer
E021
HARDWARE FAULT
System error
Contact Technical Support 1-800-225-1480
E101
CH1 TEMP OUT OF RANGE
Temperature sensor on channel 1 out of
range (5 °C to 45 °C) for ISE
Check ATC cable and probe connections. Replace if needed.
E102
CH2 TEMP OUT OF RANGE
Temperature sensor on channel 2 out of
range (5 °C to 45 °C) for ISE
Check ATC cable and probe connections. Replace if needed.
E111
CH1 CAL DUE
User calibration required on channel 1.
No calibration has been performed in
user specified time interval.
Recalibrate the analyzer.
E112
CH2 CAL DUE
User calibration required on channel 2.
No calibration has been performed in
user specified time interval.
Recalibrate the analyzer.
E127
CH1 SOUT mA OVER RANGE
Measured sensor value above Sout high
limit value on channel 1
Verify user selectable Sout high limit value in setup mode
E128
CH1 SOUT mA UNDER RANGE Measured sensor value below Sout low
limit value on channel 1
Verify user selectable Sout low limit value in setup mode
E129
CH1 TOUT mA OVER RANGE
Measured temperature value above Tout
high limit value on channel 1
Verify user selectable Tout high limit value in setup mode
E130
CH1 TOUT mA UNDER RANGE Measured temperature value below Tout
low limit value on channel 1
Verify user selectable Tout low limit value in setup mode
E131
CH2 SOUT mA OVER RANGE
Measured sensor value above Sout high
limit value on channel 2
Verify user selectable Sout high limit value in setup mode
E132
CH2 SOUT mA UNDER RANGE Measured sensor value below Sout low
limit value on channel 2
Verify user selectable Sout low limit value in setup mode
E133
CH2 TOUT mA OVER RANGE
Measured temperature value above Tout
high limit value on channel 2
Verify user selectable Tout high limit value in setup mode
E134
CH2 TOUT mA UNDER RANGE Measured temperature value below Tout
low limit value on channel 2
Verify user selectable Tout low limit value in setup mode