Analyzer Operation
III-36
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
1. Restore sample flow to the analyzer. If necessary, adjust the pressure and
flow rate through the analyzer to 40 mL/min.
2. Remove the protective cap from the reference electrode. Re-install the
electrodes on the analyzer. Be careful not to disconnect the reservoir
tubing from the reference electrode sidearm.
3. Allow the analyzer to stabilize for approximately 1 hour and then
recalibrate the system. Refer to
Chapter IV, Performing a DKA
Calibration
.
Start-Up