Thermo Scientific Medifuge Centrifuge
53
2. Troubleshooting by Guide
NOTICE
If problems occur not stated here, customer service must be
contacted�
For error numbers shown that are not described in detail in the table, follow this procedure:
1�
Restart the centrifuge�
2�
If the error message shows again, contact the customer service�
Error number
Description
Troubleshooting
E-24
Centrifuge door cannot be
opened
Check, if the centrifuge door is closed
correctly�
Restart the centrifuge�
If the error message shows again, contact
the customer service�
E-29
Drive not starting
Make sure that no objects are within the
rotor chamber�
Make sure that the motor can spin by
turning the drive with your hands�
Restart the centrifuge�
If the error message shows again, contact
the customer service�
E-31
Drive temperature high
CAUTION
Hot metal parts!
Check, if the centrifuge is accessible�
Be sure, that the room temperature is
within the limits�
Remove the rotor�
Let the centrifuge cool down for 30
minutes�
If the error message shows again, contact
the customer service�