
Thermo Scientific Medifuge Centrifuge
28
IV. Operation
1. Switching on the Centrifuge
Turn on the power switch on the back side of the centrifuge�
The centrifuge door opens automatically, if the centrifuge was closed when when switched on�
2. Open the Centrifuge Door
CAUTION
Open the centrifuge only when the rotor stopped spinning. The
display shows the current speed also during a failure.
In case of a power cut the time until the rotor is in standstill is at
least 5 min.
Never reach into the rotor chamber while the rotor is spinning.
The centrifuge door can only be opened when the centrifuge is switched on�
Press the OPEN key on the control panel�
If an error occurs, i�e� during a power failure, it is possible to open the centrifuge door using the