Thermo Scientific Medifuge Centrifuge
12
I. Technical Specifications
1. Technical Data
Thermo Scientific Medifuge Centrifuge
Environmental Conditions
For interior use
Altitudes of up to 3 000 m above sea level
Max� relative humidity 80% up to 31 °C;
decreasing linearly to 50% relative humidity at 40 °C
Environmental Conditions during
Storage and Shipping
Temperature: -10 °C to +55 °C
Humidity: 15% to 85%
Permissible Ambient Temperature
during Operation
+2 °C to +40 °C
Heat Dissipation
0�123 kWh; 419�7 Btu/h; 442�8 kJ/h
Overvoltage Category
II
Pollution Degree
2
IP
20
Running Time
99 min; hold
Maximum Speed n
max
4 900 rpm
Minimum Speed n
min
300 rpm
Maximum RCF Value at n
max
Fixed Angle Setup
Swing Out Setup
3 114 x g
3 490 x g
Noise Level at Maximum Speed
1
< 56 dB (A)
Maximum Kinetic Energy
680 J
Dimensions
Height (open door / closed door)
Width
Depth
510 mm / 240 mm
325 mm
450 mm
Weight
2
15�5 kg
1
Front Side Measurement, 1 m in front of the instrument at 1.6 m height.
2
Without Rotor.