8. Use the up and down arrow keys to adjust the temperature on LED
readout to match the temperature reading on the thermocouple meter.
9. Press the HEAT OFF button. The beeper will sound indicating that
the new calibration value just added is now stored in the nonvolatile
memory of the temperature controller.
10. Press HEAT ON button twice to complete return to normal operating
mode.
Note
It is important to press the HEAT OFF button to exit Calibration
mode.
1. Press TIMER/ELAPSED switch until TIMER and SET TIME lights
are illuminated. The HOURS or MINUTES light will also light up at
this point, depending on which option was previously chosen.
2. Press HOURS/MINUTES switch for desired timing mode.
3. Hold down appropriate arrow switch in the TIME module of the
control panel, up or down, until desired timing cycle is set from 0.1
hour up to 999 hours, or from 0.1 minute to 999 minutes depending
on which timing mode is chosen. SET TIME light will illuminate.
4. Press START to begin timed shaking; countdown will begin from time
set. TIMER and MINUTES or HOURS lights will illuminate and
timer will count down from time selected. An audible alarm will sound
at the end of the timing cycle and platform rotation will cease.
1. Press TIMER/ELAPSED membrane switch until ELAPSED light is
illuminated. The HOURS or MINUTES light will also light up at this
point depending on which option was previously chosen.
2. Press HOURS/MINUTES membrane switch for desired timing mode.
3. TIME display should show 000. Press START to begin timed shaking.
Timer will begin to count up and will display accumulated time in
display window. Platform rotation will continue until operator presses
the Stop button. The TIME display will flash when ELAPSED time
reaches 999 minutes or 999 hours.
5-6
Incubated and Refrigerated Shakers
Thermo Scientific
Section 5
Operation
Temperature
Calibration (cont.)
Setting Timer for
Timed Shaking
Setting Timer for
Continuous Shaking