CE-EMC Protection Installation Procedure
Installation
Thermo Fisher Scientific
LevelPRO Installation Guide
E-3
Reference Figure E–3 for the steps that follow.
Figure E–3.
Mounting bracket installed and wiring run through proper conduit in
detector housing
1.
Mount the bracket assembly to the detector housing. Make sure the flat
mounting plate for the EMC board housing is on the same side of the
detector housing that has the conduit fitting that is perpendicular to
the mounting tabs of the detector housing.
2.
Remove the lid of the detector housing and run the EMC protection
board wires through the conduit fitting on the housing.
3.
Apply the thread sealant supplied with the kit to the fitting threads of
the EMC assembly.
Caution
The thread sealant used must be the same as what is supplied
with the kit.
Installation