Connecting a Thermo Scientific LC System
Making Contact Closure with a Surveyor Plus LC System
36
LTQ Series Getting Connected Guide
Thermo Scientific
3. To connect a Surveyor MS Pump or Surveyor MS Pump Plus to the LC/MS interconnect
cable:
a. Connect the end of the adapter cable labeled PUMP to one of the PUMP connectors
of the 7-connector LC/MS interconnect cable.
b. Plug the end of the adapter cable labeled MS PUMP into the 8-pin socket on the rear
panel of the Surveyor MS Pump or Surveyor MS Pump Plus.
Figure 23.
Surveyor Plus LC system connected to an LTQ Series MS detector
+ -
Start In
DETECT
OR
DETECT
O
R
DETECT
OR
PUMP
PUMP
A/S
PUMP
MS PUMP
MS Detector
Autosampler
Pump
Detector
Detector
7-connector LC/MS interconnect cable
Pump
Detector
Solvent platform
Autosampler
(rear panel)
Surveyor
LC Pump
(rear panel)
Detector
(rear panel)
LTQ Series MS detector
(right-side panel)
LTQ/LXQ
interconnect
adapter cable
Not drawn to scale
Surveyor
MS Pump
(rear panel)
Adapter cable for Surveyor MS Pump
7 in. (18 cm) P/N 60053-63038
To
LC/MS interconnect
cable connector
labeled PUMP
To
Surveyor MS Pump
Connector labeled PUMP
Connector labeled PUMP
Connector
labeled
MS PUMP