Installation Manual v3.1
Thermo Fisher Corporation
Optic nitrogen purge (Optional)
Regulator valve in wall 0 – 10 bars (0 – 100 Psi)
Yes
Gas union dimension at the exit of the regulator valve
6 mm (1/4 inch
adapter supplied
with instrument)
Maximum distance between the instrument and the regulator
valve
2 m
Purity of nitrogen (Gas or Liquid)
99,995% 0
2
<10
ppm H
2
0<10
ppm
Pressure required at the input of spectrometer
6 bar (87 Psi)
Waste
Waste container, a minimum of 5 l for liquids waste
Yes
Auto Sampler rinse and waste
Minimum 1l container for the rinse supply station and waste 2 needed
Chiller (if not bought with the instrument)
Maximum distance from the spectrometer
3 m
Pipe supply O.D. dimension
12 mm
Cooling capability
900 W
Flow rate
5 l/min
Chiller inhibitor
Yes
PC and printer (if not bought with the machine)
512 Mb RAM, 40 Gb HDD, 2 Ethernet ports, 1 serial port,
USB ports, monitor, ink jet printer
Yes
Windows XP Professional (Service pack 2)
Yes