iCAP 6000 Series Pre
Chapter 6
Pre-Installation Checklist
This checklist is to be filled in by the customer to confirm that all relevant factors
concerned with the installation have been considered and dealt with.
On completion of the list, a copy should be forwarded to the local Thermo Fisher
organisation with a request for installation to be carried out. Failure to send the checklist
may result in delays with the instrument installation, any delays may incur extra charge.
Required
Completed
Location
Table dimensions for the spectrometer
900 x 600 mm
Table dimension for the auto sampler (right side of the
machine)
600 x 550 mm
Extraction
Tube diameter for extraction
125 mm
Extraction capability
10 m/s (Duo)
5 m/s (Radial)
Adjustable valve in the extraction tube
Yes
Electrical supply
Plug 200-240 VAC, 32A single phase 4000W, 50Hz
Yes
Standard mains plug 200-240 VAC for each accessory
Yes
Argon Gas
Regulator valve in the wall 0 -10 bars (0 – 100 PSI)
Yes
Gas union dimension at the exit of the regulator valve
6 mm (1/4 inch
adapter supplied
with instrument)
Maximum distance between the instrument and the regulator
valve
2 m
Purity of argon (Gas or Liquid)
99,995% 0
2
<10
ppm H
2
0<10
ppm
Pressure required at the input of spectrometer
6 bars (87 Psi)