Chapter 12
|
16-pin Communication Cable
Thermo Scientific
Single-Use System integrator’s Guide
|
64
16-pin communication
cable
This section presents the entire overview and pinout of the 16-pin
communication cable. The function of each pin is described.
Table 12.1. Pinout for 16-pin communication cable
Item
Description
Shell
Insert
Integrator connector (male
receptacle)
SH616
710216
Integrator connector
Flying leads
Flying leads
Connector manufacturer
Walther
Walther
Recommended AWG
20
20
Connector Wiring
1
Load cell communication 4-20 mA, AB RIO
Blue, Profibus RTS, or Device Net V-
2
Load cell communication Analog -4-20 mA, AB RIO
Shield, Profibus TDX.RXD+, or Device Net Can-L
3
Load cell communication AB RIO clear, Profibus TXD/
RXD+, DeviceNet drain, Modbus green
4
VFD speed control common 0 VDC
5
VFD speed control input 0-10 VDC positive
6
VFD speed control input 4-20 mA positive
7
VFD speed reference output 0-10 VDC positive
8
E-stop status dry contact or load cell communication
9
RTD output (PT-100)
10
RTD output signal - (PT-100)
11
RTD output (PT-100)
12
Temp output 4-20 mA
13
Temp output signal - 4-20 mA
14
Pressure output 4-20 mA
15
Pressure output signal - 4-20 mA
16
E-stop status dry contact or load cell communication