The instrument is very heavy. While the instrument rolls down the ramp, it must be carefully
controlled.
•
Now the HM525 NX can be rolled to its site of installation.
•
For possible return shipments, please keep all packing materials.
•
In the future, if the instrument is transported by forklift, truck, train, ship or aircraft, it must be
packed in the original shipping container with all transportation locks in place.
When Unpacking the Instrument, be sure that:
•
Enough ventilation for the cooling system is
guaranteed.
•
The distance between the wall and rear panel
is approx. 15 cm. For this, install the distance
bolts (fig 2) (item 1).
•
The suction vent areas on either side are kept
free (fig. 2) (item 2).
•
The mains switch for separating the
instrument from the power supply is accessible
at any time.
Fig 2
The installation site must also be free from:
•
Draught by open doors or by air conditioning
systems.
•
Direct exposure to sunlight into the cooling
chamber.
Note
Both measures reduce the formation of frost and
therefore result in more favourable work conditions.
High air moisture as well as high ambient temperatures
reduce the maximum performance of the instrument.
To fix the complete unit, tighten the screws (fig 2a)
(item 1) by using the attached tool.
Install the separate, packed handwheel handle.
Fig 2a