4. Enter the concentrations for all your standards; navigate to
and press
.
Follow the directions displayed in the message bar.
5. Place a cuvette with a blank solution in the measurement position and press
.
6. Present and measure your standards in order as instructed in the message bar.
After you measure the last standard, the focus will move to
. Press
to
accept the calibration data.
The screen moves to the Quant working screen.
Follow the instructions in the message bar to measure your samples.
The application supports a data table with entries for up to ten samples. Samples are labeled
S1 through S10. Once the data table is full you must clear it before you can measure
additional samples.
Tip
If you wish to re-measure a standard, navigate back to the Conc field for that
standard and follow the directions in the message bar.
Tip
You can save your method at any time by navigating to and selecting
. If you
have measured all the standards listed in the table you may save the method as
calibrated or uncalibrated. If any of the standards has not yet been measured you can
only save the method as an uncalibrated method.
Tip
Select
to go back to the calibration screen and re-measure a standard if the
graph shows that your data does not represent a good enough fit.