12
GENESYS 30 User Guide
Thermo Scientific
To analyze and store scan data
1. After the scan completes, the focus will move to the cursor tool
.
2. Use the
or
to move the cursor line left and right.
Press once for a 1 nm increment, or hold to scroll.
The cursor position and corresponding ABS/%T value are displayed above the spectrum
plot.
3. Use
to zoom in or out.
When zooming in, the screen centers on the cursor line.
4. Select
to clear the plot area.
To save and print
Data files can only be saved to a USB memory device. The on-board library is exclusively for
Method files. Data files are named as Scan_<time and date> for easy identification and
include method information and a time and date stamp.
1. Select
to save your data to a file on a USB memory device.
Data is saved as text files that can be opened by any computer spreadsheet or
word-processing application. Method files may be saved to the on-board Library or a
USB memory device. You can specify your own filename for a method.
For detailed information see the section on saving data and methods.
2. If your instrument is equipped with a printer accessory, you can print your results.
Select
to print your data to the printer.
Printed data includes a time and date stamp.
See
for detailed information on installing and loading the optional
printer accessory.
3. Select
to clear the scan plot.
You will be prompted to save your data before it is discarded.