Isotemp Freezer
Alarms | 13
Alarms
High Temp Alert
If the temperature increases above freezing your samples,
the LED light at the bottom of the cabinet will flash and the
alarm will sound. The alarm can be reset by opening and then
closing the door. The LED will continue to flash until the
cabinet temperature is below 23°F (–5°C). If the internal
temperature remains high for 15 minutes and there has not
been any power failure, contact customer care for assistance.
Door Ajar Alert
This alert will sound to alert you that the door has been open
for 5 minutes or more. The audible alert will sound until the
door is closed.
Power On Light
The green LED Power On light indicates that the unit is
properly connected to electrical power. The light glows even
when the temperature control is turned to OFF.