3
Tuning
Tuning the Reagent Ion Optics
Thermo Scientific
MS/ETD System Getting Started Guide
21
8. Select the
Reagent Ion from Vial 1
or
Reagent Ion from Vial 2
check box, whichever
appears and is available.
9. Click
Start
.
A plot of the reagent ion intensity appears (
). You can observe:
• The response of the reagent ion intensity to changes in the lens parameters in the
Reagent Ion Optics dialog box (
• The emission current, CI gas pressure, and electron energy in the Reagent Ion Source
dialog box (
)
Adjust these parameters to achieve the maximum reagent ion signal intensity.
Figure 13.
Graph view of the reagent ion source tuning in the Tune Plus window for the LTQ XL