Oring ID22XCS3 

Ring

Oring ID21XCS2 

Sensor 

Small Tab

Insert Sensor Module

Turn anti-clockwise to release

7

Elite CTS Tester Operating Instructions

 
Return of Items

Authorization must be obtained from your distributor before returning items for any reason. When applying for 
authorization, please include information regarding the reason the item(s) are to be returned.

We reserve the right to make improvements in design, construction and appearance of products without notice. 
Prices are subject to change without notice.

Sensor Replacement

Replace the sensor module at the fraction of the cost of a new tester. When the tester fails to calibrate or gives 

fluctuating readings in calibration standards, you need to change the electrode.

1.  With dry hands, grip the ring with sensor facing you. Twist the ring clockwise. Save the ring for later use.
2.  Pull the old sensor module away from the tester.
3.  Align the four tabs on the new module so that they match the four slots on the tester.
4.  Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto the new sensor  

module. Push the other O-ring over the module and thread it into place by firmly twisting counter clockwise.

 

Note:

 It is necessary that you recalibrate your tester prior to measurement after a sensor  replacement.

Self Diagnostic Messages

      

Batteries are weak and need replacement soon.

stable error

Appears when calibration is attempted but the reading is not yet stable. Wait for the 

reading to stabilize or manually confirm the calibration by pressing enter.

over range

The reading is above the measuring range of tester.

Summary of Contents for Elite CTS

Page 1: ...Elite CTS Tester Operating Instructions 68X001308 Rev 0 June 2017...

Page 2: ...erating Instructions This instruction manual is applicable for Elite CTS Tester Replacing the Batteries The Elite CTS tester uses four AAA 1 5 V batteries 1 To remove the battery cover See Figure 1 Cl...

Page 3: ...Press to enter setup window Scroll down by pressing to select Temp Set Press to select Temp set The display shows Set C F Temp Cal and Temp Coeff 2 Press to select Set C F Scroll down by pressing to t...

Page 4: ...red off The conductivity value can be calibrated automatically or manually while the TDS salinity values require manual calibration The tester will begin in the measurement mode that was used when it...

Page 5: ...imer stops blinking and will appear to indicate the stability of the reading CAUTION Testing dry samples is not accurate and can lead to sensor damage or breakage Soils must be wet and free of particu...

Page 6: ...its protective cap For cup type electrodes remove the white plastic cup and insert to thoroughly clean viscous solutions Never scratch electrodes with a hard substance 2 For better performance soak th...

Page 7: ...eed to change the electrode 1 With dry hands grip the ring with sensor facing you Twist the ring clockwise Save the ring for later use 2 Pull the old sensor module away from the tester 3 Align the fou...

Page 8: ...to 1000 ppm 0 10 to 10 00 ppt TDS Factor 0 5 Resolution 0 1 ppm 1 ppm 0 01 ppt Relative accuracy 1 full scale Calibration points Up to 3 points TDS Factor 0 40 to 1 00 Selectable Salinity Salinity ra...

Page 9: ...TSPIN Conductivity TDS salinity pocket tester with batteries ELITECTSCUP Conductivity TDS salinity pocket tester with batteries Conductivity Standard Solutions and Sachets ECCON100BT 100 S cm KCl 480...

Reviews: