Elite CTS Tester Operating Instructions

5.  Dip the tester into a solution of known temperature and allow time for the in built temperature sensor to stabilize.

6.  Press      to adjust the temperature value or press the 

 to confirm the calibrated value as the new temperature 

value of the solution.

Note: 

To exit this p

rogram without confirming the calibration, press 

 
 

Temp Coefficient

1.  Press 

 to enter setup window. Scroll down by pressing 

 to select Temp Set. 

2.  Press 

 to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.

3.  Scroll down by pressing      to toggle between Set °C/°F, Temp Cal and Temp Coeff.

4.  Press 

 to select Temp Coeff or 

 to adjust the Temp Coeff.

5.  Press 

 to confirm the 

Temp Coeff value. 

the new value is automatically confirmed with a 

 .

Calibration for Conductivity, TDS, or Salinity

For best results, periodic calibration with an accurate standard is recommended prior to measurement. Use the 
calibration standard value that is close to your intended sample value.The tester will retain one calibration value in 
each mode (conductivity, TDS, salinity) when the instrument is powered off. The conductivity value can be calibrated 
automatically or manually, while the TDS & salinity values require manual calibration. The tester will begin in the 
measurement mode that was used when it was powered off. See “Elite CTS Measurement Parameter Setting” to 
change the desired parameter.

Automatic Calibration for Conductivity

1.  Remove the cap and press the       to power on.
2.  Dip the sensor in at least 30 mm of calibration standard.
3.  Stir gently and press 

 to begin the calibration.

4.  The display will show CAL followed by the default value. CAL is indicated on the display during calibration mode.
5.  If the reading is within the calibration range of the automatically recognized standards; 80 (84 µS/cm),  

1410 (1413 µS/cm), or 12.90 (12.88 mS/cm), the   icon is displayed when the automatic calibration standard 
value has been detected.

6.  Press 

 to accept the auto conductivity standard and finish the calibration. 

7.  Display returns to Measurement window.

 

Measure

Settings 

 

Reset

Key Info   

Scroll 

Accept 

EXIT 

Set °C/°F
Temp Cal

Adjust  

Temp Coeff 

2.00

2.00

Temp Set

Temp Coeff

Measure

Settings 

 

Reset

Key Info   

Scroll 

Accept 

EXIT 

Set °C/°F

Temp Coeff

Temp Cal 
23.4 °C 
23.4 °C

23.4°C

Temp Set

Temp Cal

 

CAL

µS/cm

 844

1413

Place the tester 
in the calibration 
STD to start 
or press    to 
escape.

 

COND

µS/cm

 844

23.5 °C

CAL

Summary of Contents for Elite CTS

Page 1: ...Elite CTS Tester Operating Instructions 68X001308 Rev 0 June 2017...

Page 2: ...erating Instructions This instruction manual is applicable for Elite CTS Tester Replacing the Batteries The Elite CTS tester uses four AAA 1 5 V batteries 1 To remove the battery cover See Figure 1 Cl...

Page 3: ...Press to enter setup window Scroll down by pressing to select Temp Set Press to select Temp set The display shows Set C F Temp Cal and Temp Coeff 2 Press to select Set C F Scroll down by pressing to t...

Page 4: ...red off The conductivity value can be calibrated automatically or manually while the TDS salinity values require manual calibration The tester will begin in the measurement mode that was used when it...

Page 5: ...imer stops blinking and will appear to indicate the stability of the reading CAUTION Testing dry samples is not accurate and can lead to sensor damage or breakage Soils must be wet and free of particu...

Page 6: ...its protective cap For cup type electrodes remove the white plastic cup and insert to thoroughly clean viscous solutions Never scratch electrodes with a hard substance 2 For better performance soak th...

Page 7: ...eed to change the electrode 1 With dry hands grip the ring with sensor facing you Twist the ring clockwise Save the ring for later use 2 Pull the old sensor module away from the tester 3 Align the fou...

Page 8: ...to 1000 ppm 0 10 to 10 00 ppt TDS Factor 0 5 Resolution 0 1 ppm 1 ppm 0 01 ppt Relative accuracy 1 full scale Calibration points Up to 3 points TDS Factor 0 40 to 1 00 Selectable Salinity Salinity ra...

Page 9: ...TSPIN Conductivity TDS salinity pocket tester with batteries ELITECTSCUP Conductivity TDS salinity pocket tester with batteries Conductivity Standard Solutions and Sachets ECCON100BT 100 S cm KCl 480...

Reviews: