Thermo Scientific Elite CTS Operating Instructions Manual Download Page 9

9

Elite CTS Tester Operating Instructions

Specification

Elite CTS

General

Display

Graphics, Dot Matrix 80X100 Pixel

Auto off

8.5 minutes ( From Last Key press)

Reset

User / Factory

Power requirement

Four AAA 1.5 V batteries

Battey life

>150 hours

Water proofing

IP67

Environmental operating conditions

Ambient operating temperature

5°C to 45°C 

Relative humidity

5% to 85% Non-Condensing

Storage temperature

-20°C to 60°C 

Storage humidity

5% to 85% Non-Condensing

Accessories

Ordering Code

Product Description

CTS Tester

ELITECTSPIN

Conductivity/TDS/salinity pocket tester with batteries

ELITECTSCUP

Conductivity/TDS/salinity pocket tester with batteries

Conductivity Standard Solutions and Sachets

ECCON100BT

100 µS/cm KCl, 480 mL

ECCON1413BT

1413 µS/cm KCl, 480 mL

ECCON1288BT

12.88 mS/cm KCl, 480 mL

ECCON1413BS

1413 µS/cm KCl, box of 20 x 20 mL sachets

TDS 442* Standard Solutions

EC442300BT

300 ppm 442, 480 mL

EC4421000BT

1000 ppm 442, 480 mL

Salinity (NaCl) Solutions

ECNACL5PPT

5 ppt NaCl, 480 mL

Other Accessories

ELITECAP

Replacement sensor cap

TFLANYARD

Tester lanyard

*The 442 scale simulates fresh water dissolved solids as a mixture of 4 parts sodium sulfate, 4 parts sodium  
bicarbonate, and 2 parts sodium chloride (40% Na2SO4,40% NaHCO3, 20% NaCl).

Summary of Contents for Elite CTS

Page 1: ...Elite CTS Tester Operating Instructions 68X001308 Rev 0 June 2017...

Page 2: ...erating Instructions This instruction manual is applicable for Elite CTS Tester Replacing the Batteries The Elite CTS tester uses four AAA 1 5 V batteries 1 To remove the battery cover See Figure 1 Cl...

Page 3: ...Press to enter setup window Scroll down by pressing to select Temp Set Press to select Temp set The display shows Set C F Temp Cal and Temp Coeff 2 Press to select Set C F Scroll down by pressing to t...

Page 4: ...red off The conductivity value can be calibrated automatically or manually while the TDS salinity values require manual calibration The tester will begin in the measurement mode that was used when it...

Page 5: ...imer stops blinking and will appear to indicate the stability of the reading CAUTION Testing dry samples is not accurate and can lead to sensor damage or breakage Soils must be wet and free of particu...

Page 6: ...its protective cap For cup type electrodes remove the white plastic cup and insert to thoroughly clean viscous solutions Never scratch electrodes with a hard substance 2 For better performance soak th...

Page 7: ...eed to change the electrode 1 With dry hands grip the ring with sensor facing you Twist the ring clockwise Save the ring for later use 2 Pull the old sensor module away from the tester 3 Align the fou...

Page 8: ...to 1000 ppm 0 10 to 10 00 ppt TDS Factor 0 5 Resolution 0 1 ppm 1 ppm 0 01 ppt Relative accuracy 1 full scale Calibration points Up to 3 points TDS Factor 0 40 to 1 00 Selectable Salinity Salinity ra...

Page 9: ...TSPIN Conductivity TDS salinity pocket tester with batteries ELITECTSCUP Conductivity TDS salinity pocket tester with batteries Conductivity Standard Solutions and Sachets ECCON100BT 100 S cm KCl 480...

Reviews: