 Page 

5  

Getting Started 

Select a location that allows for 3" clearance behind the power supply, and comfortable reach of 
the front panel controls and cell connections.  Do not block the vented area of the case - on the 
front bottom of the unit, or the fan area at the rear.  Connect the power supply to a 3-prong 
grounded AC outlet, 

using the AC cord provided with the unit only

.  Connect the 

electrophoresis apparatus to the power supply, making sure to match the red positive lead to the 
red positive jack, and the black negative lead to the black negative jack.  Similarly, connect one or 
more electrophoresis apparatus to the power supply if you plan to run more than one in parallel.  
Power the unit on using the AC power switch located at the rear of the unit, next to the AC line 
cord entry.  
 

 
 
 

 

 

Figure 1. Front Panel Controls 

 

Using the Power Supply 

Press the blue key on the front of the unit to enable/disable the control logic.  The LCD display 
will illuminate and show the values of the last saved run setup.  The EC200 preserves the run 
settings each time you start a run.  One set of conditions is saved for each mode, constant voltage 
or constant current.  The setpoint value, limit parameter, time duration, and power-fail restart 
settings are all saved.  This allows convenient setup for repetitive runs. 

To change the control mode, press the Exit key.  From the resulting screen, select either constant 
volts or constant amps using the function keys to the right of the display. 

After selecting a control mode, either constant voltage or current, and setting the limit and timed 
run parameters if desired (see Constant Voltage/Current Operations) press the Run key to energize 
the power supply output.  The power supply output will ramp up to the appropriate setpoint, while 
not allowing the limit parameter to be exceeded.  If the limit parameter should be exceeded, then 
the power supply will crossover control modes, making the limit parameter the new control 

black negative jacks 

black negative jacks 

  red positive jacks 

  red positive jacks 

blue ON/OFF key 

Stop key 

Run key 

Pause key 

Function keys 

Exit key 

Clear key 

Summary of Contents for EC200XL

Page 1: ...LTM1595X2 8 28 07 2007 All rights reserved Printed in U S A EC200XL Electrophoresis Power Supply ...

Page 2: ...on 3 Unpacking the Power Supply 3 Specifications 4 Getting Started 5 Using the Power Supply 5 Constant Voltage Operations 6 Constant Current Operations 6 Timed Operations 6 Automatic Power fail Restart 7 Cleaning 7 Troubleshooting and Error Indications 8 EC200 Warranty Statement 9 Compliance 9 Replacement Parts 9 ...

Page 3: ...ated apparatus for any signs of wear or damage before using with this equipment Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current capabilities of the EC200 power supply The output of the EC200 power supply is intended for connection to electrically isolated electrophoresis apparatus only Use only with electrically isolated electrophoresis apparatus wi...

Page 4: ... If this non constant limit is reached the power supply will automatically switch control modes from constant voltage to constant current or vice versa In this way the EC200 protects your electrophoresis cells from damaging over power conditions The EC200 power supply also provides for timed operation in either voltage or current modes and allows an automatic completion in the event of a power los...

Page 5: ...0 60Hz 250VA Max Environmental Operating temperature 0 40 C 0 95 R H non condensing Altitude 2000m Overvoltage category II IEC664 Pollution degree 2 IEC664 DC output Power 5 200VDC 200W Max 0 01 2 00A 200W Max Ripple 1 Drift 1 after 30 minute warm up ...

Page 6: ...ate and show the values of the last saved run setup The EC200 preserves the run settings each time you start a run One set of conditions is saved for each mode constant voltage or constant current The setpoint value limit parameter time duration and power fail restart settings are all saved This allows convenient setup for repetitive runs To change the control mode press the Exit key From the resu...

Page 7: ...et to the power supply maximum value of 2 00 A The EC200 will not accept a current limit outside of the allowed range In addition the EC200 will never exceed the maximum power output specification of 200W When starting a run the power supply calculates a maximum limit parameter above which the power supply maximum power output rating would be exceeded If this value is lower than the limit setpoint...

Page 8: ... restart press the top function key when in the operating mode screen constant voltage or current The state of power fail restart will toggle from off to on or on to off with each keypress When AC power is restored during a timed run in which power fail restart is enabled the display will show POWER FAIL RESTART for approximately 10 seconds indicating power fail restart pending During this time th...

Page 9: ...ctions 02 Step load current change The EC200 has detected a large change in load current Check your setup and connections 04 Loss of AC power PF disabled and not completed The EC200 detected loss of AC power the run was not completed since PF was not enabled 05 Loss of AC power run not timed The EC200 detected loss of AC power 06 99 The EC200 detected an internal error condition Contact technical ...

Page 10: ...chaser s exclusive remedy for any claim arising out of the purchase of any product listed above is repair replacement or credit as described above where applicable In no event 1 shall the cost of the exclusive remedy exceed the purchase price 2 shall Thermo be liable for any special indirect incidental consequential or exemplary damages howsoever arising even if Thermo has been advised of the poss...

Reviews: