background image

 Page 

9  

EC200 Warranty Statement  

The Thermo Scientific Company ("Thermo") warrants to the direct purchaser that the EC200 will 
be free from defects in material or workmanship for a specified warranty period.  During that 
period, Thermo

 

will repair or replace the product or provide credit, at its sole option, upon prompt 

notification and compliance with its instructions.  For EC200 power supplies that specified period 
is 48 months from manufacturing date. 

Unless otherwise agreed, the warranty is limited to the country in which the product is sold. 

No Thermo

 

employee, agent or representative has the authority to bind Thermo

 

to any oral 

representation or warranty concerning any product sold.  Any oral representation or warranty 
made prior to purchase of any product and not set forth in writing and signed by a duly authorized 
officer of Thermo

 

shall not be enforceable by the purchaser. 

THERMO EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, 
INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A 
PARTICULAR PURPOSE. 

Thermo's sole responsibility and the purchaser's exclusive remedy for any claim arising out of the 
purchase of any product listed above is repair, replacement or credit as described above, where 
applicable.  In no event: 1) shall the cost of the exclusive remedy exceed the purchase price: 2) 
shall Thermo be liable for any special, indirect, incidental, consequential, or exemplary damages, 
howsoever arising, even if Thermo has been advised of the possibility of such damages. 

Each article that Thermo furnishes will conform to the written specifications given in this manual, 
or those of a further improved model.  Changes are made often to the information in the manual 
and will be incorporated into future editions. 

Compliance 

CSA 1010 

This equipment has been designed and tested to conform to CSA1010 safety standards, as 
applicable to laboratory instrumentation.  This applies only to the EC200 when used as 
specified in the documentation, in its intended applications, with Thermo Scientific 
approved electrophoresis apparatus only.  Usage in any other manner may not provide 
suitable protection. 
 

Replacement Parts 

AC line cord 

 

 

FB-CORD-1 

Fuse, 

T3A,250V 

  FB-FUSE-2 

 

For replacement parts contact Thermo Scientific at: 

 

Technical Support: 1-800-943-2006 or 1-800-926-0505 

Accessories 

Adapter for Cells 

 

FBAD-1 

Under Counter Bracket   

FBUB3000

 

Summary of Contents for EC200XL

Page 1: ...LTM1595X2 8 28 07 2007 All rights reserved Printed in U S A EC200XL Electrophoresis Power Supply ...

Page 2: ...on 3 Unpacking the Power Supply 3 Specifications 4 Getting Started 5 Using the Power Supply 5 Constant Voltage Operations 6 Constant Current Operations 6 Timed Operations 6 Automatic Power fail Restart 7 Cleaning 7 Troubleshooting and Error Indications 8 EC200 Warranty Statement 9 Compliance 9 Replacement Parts 9 ...

Page 3: ...ated apparatus for any signs of wear or damage before using with this equipment Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current capabilities of the EC200 power supply The output of the EC200 power supply is intended for connection to electrically isolated electrophoresis apparatus only Use only with electrically isolated electrophoresis apparatus wi...

Page 4: ... If this non constant limit is reached the power supply will automatically switch control modes from constant voltage to constant current or vice versa In this way the EC200 protects your electrophoresis cells from damaging over power conditions The EC200 power supply also provides for timed operation in either voltage or current modes and allows an automatic completion in the event of a power los...

Page 5: ...0 60Hz 250VA Max Environmental Operating temperature 0 40 C 0 95 R H non condensing Altitude 2000m Overvoltage category II IEC664 Pollution degree 2 IEC664 DC output Power 5 200VDC 200W Max 0 01 2 00A 200W Max Ripple 1 Drift 1 after 30 minute warm up ...

Page 6: ...ate and show the values of the last saved run setup The EC200 preserves the run settings each time you start a run One set of conditions is saved for each mode constant voltage or constant current The setpoint value limit parameter time duration and power fail restart settings are all saved This allows convenient setup for repetitive runs To change the control mode press the Exit key From the resu...

Page 7: ...et to the power supply maximum value of 2 00 A The EC200 will not accept a current limit outside of the allowed range In addition the EC200 will never exceed the maximum power output specification of 200W When starting a run the power supply calculates a maximum limit parameter above which the power supply maximum power output rating would be exceeded If this value is lower than the limit setpoint...

Page 8: ... restart press the top function key when in the operating mode screen constant voltage or current The state of power fail restart will toggle from off to on or on to off with each keypress When AC power is restored during a timed run in which power fail restart is enabled the display will show POWER FAIL RESTART for approximately 10 seconds indicating power fail restart pending During this time th...

Page 9: ...ctions 02 Step load current change The EC200 has detected a large change in load current Check your setup and connections 04 Loss of AC power PF disabled and not completed The EC200 detected loss of AC power the run was not completed since PF was not enabled 05 Loss of AC power run not timed The EC200 detected loss of AC power 06 99 The EC200 detected an internal error condition Contact technical ...

Page 10: ...chaser s exclusive remedy for any claim arising out of the purchase of any product listed above is repair replacement or credit as described above where applicable In no event 1 shall the cost of the exclusive remedy exceed the purchase price 2 shall Thermo be liable for any special indirect incidental consequential or exemplary damages howsoever arising even if Thermo has been advised of the poss...