setpoint.  The run mode display shows the controlled parameter highlighted, and will 
automatically change if a crossover has occurred. 

To stop a run in progress, press the Run key, the Stop key, or the blue ON/OFF key.  The display 
will change to "STOP" signifying that the output is no longer energized.  To pause a run in 
progress, press the Pause key.  The display will change to "PAUSE" signifying that the output is 
no longer energized, and that the run can be resumed.  Pressing Run or Pause resumes the run 
from the previous point (i.e. for a timed run).  Only a run that has been paused can be restarted.  
After Stopping a run, or to change parameters of a (paused) run, you must cycle the control logic 
off and then back on by pressing the blue ON/OFF key. 

During a run, the display will show the actual values of voltage and current.  The time value 
displays elapsed time for an untimed run, or time left for a timed run. 

Constant Voltage Operations 

From the mode selection screen, select constant voltage operation by pressing the appropriate 
function key located to the right of the display.  To get to the selection screen from an operating 
screen, press Exit, or cycle the control logic off and then back on by pressing the blue ON/OFF 
key.  The EC200 will display the last saved values for constant voltage mode.  To change a 
parameter, press the function key located to the right of the display, next to the parameter you 
wish to change.  The EC200 will allow a new value to be entered using the keypad numeric keys.  
Press the function key labeled "Enter" to complete the new setting.  Press Clear to erase previous 
keystrokes.  Press Exit to cancel entering a new value.  The EC200 will not accept a voltage 
setpoint outside of the allowed range.  Press Clear to re-enter an appropriate value. 

The limit parameter, in this case current, is normally set to the power supply maximum value of 
2.00 A.  The EC200 will not accept a current limit outside of the allowed range. 

In addition, the EC200 will never exceed the maximum power output specification of 200W.  
When starting a run, the power supply calculates a maximum limit parameter, above which the 
power supply maximum power output rating would be exceeded.  If this value is lower than the 
limit setpoint entered, the EC200 will automatically lower the limit parameter. 

Constant Current Operations 

From the mode selection screen, select constant current operation by pressing the appropriate 
function key located to the right of the display.  To get to the selection screen from an operating 
screen, press Exit, or cycle the control logic off and then back on by pressing the blue ON/OFF 
key.  The EC200 will display the last saved values for constant current mode.  To change a 
parameter, press the function key located to the right of the display, next to the parameter you 
wish to change.  The EC200 will allow a new value to be entered using the keypad numeric keys.  
Press the function key labeled "Enter" to complete the new setting.  Press Clear to erase previous 
keystrokes.  Press Exit to cancel entering a new value.  The EC200 will not accept a current 
setpoint outside of the allowed range.  Press Clear to re-enter an appropriate value. 

The limit parameter, in this case voltage, is normally set to the power supply maximum value of 
200 volts.  The EC200 will not accept a voltage limit outside of the allowed range. 

In addition, the EC200 will never exceed the maximum power output specification of 200W.  
When starting a run, the power supply calculates a maximum limit parameter, above which the 
power supply maximum power output rating would be exceeded.  If this value is lower than the 
limit setpoint entered, the EC200 will automatically lower the limit parameter. 

Timed Operations 

It is possible to enter an amount of time for the power supply to provide power, after which it will 
automatically shut off.  Use this feature for timed runs. 

Summary of Contents for EC200XL

Page 1: ...LTM1595X2 8 28 07 2007 All rights reserved Printed in U S A EC200XL Electrophoresis Power Supply ...

Page 2: ...on 3 Unpacking the Power Supply 3 Specifications 4 Getting Started 5 Using the Power Supply 5 Constant Voltage Operations 6 Constant Current Operations 6 Timed Operations 6 Automatic Power fail Restart 7 Cleaning 7 Troubleshooting and Error Indications 8 EC200 Warranty Statement 9 Compliance 9 Replacement Parts 9 ...

Page 3: ...ated apparatus for any signs of wear or damage before using with this equipment Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current capabilities of the EC200 power supply The output of the EC200 power supply is intended for connection to electrically isolated electrophoresis apparatus only Use only with electrically isolated electrophoresis apparatus wi...

Page 4: ... If this non constant limit is reached the power supply will automatically switch control modes from constant voltage to constant current or vice versa In this way the EC200 protects your electrophoresis cells from damaging over power conditions The EC200 power supply also provides for timed operation in either voltage or current modes and allows an automatic completion in the event of a power los...

Page 5: ...0 60Hz 250VA Max Environmental Operating temperature 0 40 C 0 95 R H non condensing Altitude 2000m Overvoltage category II IEC664 Pollution degree 2 IEC664 DC output Power 5 200VDC 200W Max 0 01 2 00A 200W Max Ripple 1 Drift 1 after 30 minute warm up ...

Page 6: ...ate and show the values of the last saved run setup The EC200 preserves the run settings each time you start a run One set of conditions is saved for each mode constant voltage or constant current The setpoint value limit parameter time duration and power fail restart settings are all saved This allows convenient setup for repetitive runs To change the control mode press the Exit key From the resu...

Page 7: ...et to the power supply maximum value of 2 00 A The EC200 will not accept a current limit outside of the allowed range In addition the EC200 will never exceed the maximum power output specification of 200W When starting a run the power supply calculates a maximum limit parameter above which the power supply maximum power output rating would be exceeded If this value is lower than the limit setpoint...

Page 8: ... restart press the top function key when in the operating mode screen constant voltage or current The state of power fail restart will toggle from off to on or on to off with each keypress When AC power is restored during a timed run in which power fail restart is enabled the display will show POWER FAIL RESTART for approximately 10 seconds indicating power fail restart pending During this time th...

Page 9: ...ctions 02 Step load current change The EC200 has detected a large change in load current Check your setup and connections 04 Loss of AC power PF disabled and not completed The EC200 detected loss of AC power the run was not completed since PF was not enabled 05 Loss of AC power run not timed The EC200 detected loss of AC power 06 99 The EC200 detected an internal error condition Contact technical ...

Page 10: ...chaser s exclusive remedy for any claim arising out of the purchase of any product listed above is repair replacement or credit as described above where applicable In no event 1 shall the cost of the exclusive remedy exceed the purchase price 2 shall Thermo be liable for any special indirect incidental consequential or exemplary damages howsoever arising even if Thermo has been advised of the poss...

Reviews: