Thermo Scientific DLT Installation & Operating Manual Download Page 7

Thermo Scientific Harris Ultra-Low Temperature Freezers

 Installation and Operation

3

4.6 Compressor Mounts

Some Harris DLTs are shipped with the compressor mounts 
secured. The springs under the compressor mounting feet are 
fully compressed by hold-down nuts for shipping. 

These nuts 

must be loosened before the unit is put into operation

 (refer to 

Figure 2).

To loosen the compressor mount springs, complete the following 
steps:

1. Remove the end panel on the right side of the machine 

compartment to expose the compressor.

2. Use a 3/4 in. box end or socket wrench with ratcheting handle 

and a 4 in. extension to loosen the nuts on the compressor 
mounts (refer to Figure 2). Loosen the nuts until the 
clearance between the bottom of the nut and the top of the 

mounting foot is 1/8 in.

3. Inspect the area under each mounting foot for spacer clips. 

Any clips must be removed before compressor operation.

Figure 2.  Compressor Mount Springs

Note:

Some of the Harris DLT freezers require loosening the 

compressor bolts. This is not necessary for all models.

4.7 Pressure Equalization Port

When an upright ultra-low temperature freezer door is opened, 
room temperature air rushes into the storage compartment. 
When the door is closed, the fixed volume of air is cooled rapidly. 
Pressure drops below atmospheric pressure, resulting in a 

substantial vacuum. Re-entry into the cabinet is impossible until 
internal pressures are returned to atmospheric pressure. Without a 
pressure equalization mechanism, it can take several minutes 

before the door can easily be reopened. 

All upright models feature a port that provides vacuum relief 
after door openings. 

The pressure equalization port is located in the bottom left side of 
the storage chamber. Although the port is designed to self-
defrost, excessive frost accumulation in the bottom of the 

chamber could eventually restrict air flow. Therefore you should 
periodically inspect the port and brush away any loose frost using 
a stiff nylon brush.

WARNING!

To ensure proper operation, be sure not 

to obstruct the port. Push any storage racks near the 
port all the way to the rear of the cabinet.

WARNING!

When the freezer is in operation, avoid 

touching the lower left side of the cabinet exterior near 
the “hot surface” warning label. The port can get hot 
immediately after it is activated. 

Spring

Summary of Contents for DLT

Page 1: ...Installation and Operation Manual Thermo Scientific Harris DLT SLT and ELT Ultra Low Temperature Freezers...

Page 2: ...d trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights reserved Suva is a regist...

Page 3: ...ocking the Door 2 4 5 2 Unlocking the Door 2 4 5 3 Opening the Door 2 4 6 Compressor Mounts 3 4 7 Pressure Equalization Port 3 5 Operation Harris DLT and SLT Series Freezers 4 5 1 Control Panel Featur...

Page 4: ...arm 9 8 Maintenance and Troubleshooting 10 8 1 Condenser Maintenance 10 8 1 1 Cleaning the Condenser 10 8 1 2 Cleaning the Condenser Filter 10 8 2 Gasket Maintenance 10 8 3 Defrosting the Freezer 10 8...

Page 5: ...the compressors to operate for a prolonged time period Attempting to utilize this freezer improperly may jeopardize user product safety or cause undue stress or damage to the refrigeration compressors...

Page 6: ...the prong is removed the warranty is invalidated 4 3 Voltage Safeguard Freezers are equipped with a standard built in Voltage Safeguard circuit to detect and adjust low line voltage The Voltage Safeg...

Page 7: ...qualization Port When an upright ultra low temperature freezer door is opened room temperature air rushes into the storage compartment When the door is closed the fixed volume of air is cooled rapidly...

Page 8: ...rease temperature values 18 Increment Pad Use this pad to increase temperature values The following indicators are available only when the freezer is equipped with the optional backup system refer to...

Page 9: ...the Warm Alarm Set Point 4 Release both pads when the digital temperature display window shows the correct set point value When the cabinet temperature drops below the Warm Alarm setting turn the key...

Page 10: ...serve the following precautions when using liquid CO2 and LN2 backup systems WARNING If a CO2 or LN2 cylinder falls and a valve is knocked off the cylinder becomes a deadly and completely unguided mis...

Page 11: ...wer WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 4 System Status Lights In addition to the backup control system the built in backup system includ...

Page 12: ...ty 5 6 6 5 7 5 14 ft3 Empty 5 4 6 2 7 0 1 2 Full 4 8 5 5 6 3 1 2 Full 4 5 5 2 5 9 Full 3 5 4 2 4 8 Full 3 4 3 9 4 4 17 ft3 Empty 6 5 7 5 8 7 17 ft3 Empty 6 2 7 1 8 1 1 2 Full 5 5 6 3 7 3 1 2 Full 5 3...

Page 13: ...t Point and simultaneously press and hold the Set Point Display button The temperature display changes to read the existing set point value 2 Turn the set point screw clockwise for a colder setting an...

Page 14: ...inet 2 Turn off the freezer 3 Open the outer door and all inner doors 4 Let the freezer stand with doors open for at least 24 hours This allows both the interior and foamed refrigerant system to warm...

Page 15: ...a virtually unlimited standby life unless they develop a leak These containers hold liquid at room temperature and high pressure Several bottles can be connected together increasing the effective len...

Page 16: ...oxes for all chest models except 3 1 and 4 5 cu ft 6230 Single split rack holds 4 3 in boxes for all chest models except 3 1 and 4 5 cu ft 6231 Basket Rail Kit for 6 8 cu ft chest freezers 6217 Basket...

Page 17: ...rogen cylinder Description Catalog No Temperature recorder 6 in circular chart seven day drive panel mounted for 40 C to 95 C Not available for Harris Classic chest models which use 6383 6 Factory ins...

Page 18: ...rdering 5612 Deluxe Electronic Remote Alarm System User programmable to sound alarm in the event of temperature rise or power failure Can dial up to four telephone numbers to advise of alarm condition...

Page 19: ...die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi...

Page 20: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 21: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 38257H01 Rev J...

Reviews: