Thermo Scientific DLT Installation & Operating Manual Download Page 14

Thermo Scientific Harris Ultra-Low Temperature Freezers

 Installation and Operation

10

8 Maintenance and Troubleshooting

WARNING!

Unauthorized repair of your freezer will 

invalidate your warranty. Contact Technical Service at 
1-800-438-4851 for additional information.

CAUTION!

Maintenance should only be performed 

by trained personnel.

8.1 Condenser Maintenance

8.1.1 Cleaning the Condenser

Clean the condenser at least every six months; more often if the 

laboratory area is extremely dust prone. 

To clean the condenser, complete the following steps:

1. Pull the grill open.

2. Remove the filter. Check the fans. If a fan is not operating, 

contact an Authorized Service Company immediately.

3. Vacuum the condenser.

4. Replace the filter and close the grill.

8.1.2 Cleaning the Condenser Filter

Clean the condenser filter every two or three months.

1. Pull the grill open.
2. Remove the filter.
3. Shake the filter to remove loose dust, rinse the filter in clean 

water, shake the excess water from the filter, and replace the 
filter.

4. Close the grill.

8.2 Gasket Maintenance

Periodically check the gaskets around the door or lid for 
punctures or tears. Leaks are indicated by a streak of frost which 

forms at the point of gasket failure. Make sure that the cabinet is 
level (refer to Section 4.4 on page 2 for leveling information).

Keep the lid and door gaskets clean and frost free by wiping 

gently with a soft cloth.

8.3 Defrosting the Freezer

Defrost the freezer once or year or whenever the ice buildup 
exceeds 3/8”.

To defrost, complete the following steps:

1. Remove all products and place in another cabinet.

2. Turn off the freezer.
3. Open the outer door and all inner doors.
4. Let the freezer stand with doors open for at least 24 hours. 

This allows both the interior and foamed refrigerant system 
to warm to room temperature.

5. Dispose of the ice and wipe out any water standing in the 

bottom of the cabinet.

6. If there is freezer odor, wash the interior with a solution of 

baking soda and warm water. Clean the exterior with any 

common household cleaning wax.

7. Close the doors, restart the freezer and reload, following the 

instructions in Section 2.1 on page 1.

8.4 Alarm Battery Maintenance

Have a technician check the condition of the alarm battery at 
least once a year.

To replace the alarm battery, complete the following steps:

1. Remove the front grill. The alarm battery is located directly 

behind the grill. The terminals are the “push on” type.

2. Grasp the terminal with pliers and work it gently back and 

forth while pulling it off. The fittings are tight.

3. Remove the battery and put the new battery in place.

Note:

You may have to cut a strip of silicone rubber in order to 
remove the battery.

4. Connect the battery terminals and replace the front grill.

8.5 Troubleshooting Procedures

WARNING!

Troubleshooting procedures involve 

working with high voltages which can cause injury or 
death. Troubleshooting should only be performed by 
trained personnel.

Table 2. 

Troubleshooting Procedures

Problem

Solution

Clean Filter 
indicator is on.

Condenser filter is dirty. Refer to Section 8.1.2.

Alarm Battery Low 
indicator is on.

Alarm battery is low on charge. Have a 
technician check the alarm battery. Refer to 
Section 8.4.

Voltage Low 
indicator is on.

Line voltage is low and the freezer has activated 
the Voltage boost system to compensate. Call an 
electronics technician to check the power supply.

Summary of Contents for DLT

Page 1: ...Installation and Operation Manual Thermo Scientific Harris DLT SLT and ELT Ultra Low Temperature Freezers...

Page 2: ...d trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights reserved Suva is a regist...

Page 3: ...ocking the Door 2 4 5 2 Unlocking the Door 2 4 5 3 Opening the Door 2 4 6 Compressor Mounts 3 4 7 Pressure Equalization Port 3 5 Operation Harris DLT and SLT Series Freezers 4 5 1 Control Panel Featur...

Page 4: ...arm 9 8 Maintenance and Troubleshooting 10 8 1 Condenser Maintenance 10 8 1 1 Cleaning the Condenser 10 8 1 2 Cleaning the Condenser Filter 10 8 2 Gasket Maintenance 10 8 3 Defrosting the Freezer 10 8...

Page 5: ...the compressors to operate for a prolonged time period Attempting to utilize this freezer improperly may jeopardize user product safety or cause undue stress or damage to the refrigeration compressors...

Page 6: ...the prong is removed the warranty is invalidated 4 3 Voltage Safeguard Freezers are equipped with a standard built in Voltage Safeguard circuit to detect and adjust low line voltage The Voltage Safeg...

Page 7: ...qualization Port When an upright ultra low temperature freezer door is opened room temperature air rushes into the storage compartment When the door is closed the fixed volume of air is cooled rapidly...

Page 8: ...rease temperature values 18 Increment Pad Use this pad to increase temperature values The following indicators are available only when the freezer is equipped with the optional backup system refer to...

Page 9: ...the Warm Alarm Set Point 4 Release both pads when the digital temperature display window shows the correct set point value When the cabinet temperature drops below the Warm Alarm setting turn the key...

Page 10: ...serve the following precautions when using liquid CO2 and LN2 backup systems WARNING If a CO2 or LN2 cylinder falls and a valve is knocked off the cylinder becomes a deadly and completely unguided mis...

Page 11: ...wer WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 4 System Status Lights In addition to the backup control system the built in backup system includ...

Page 12: ...ty 5 6 6 5 7 5 14 ft3 Empty 5 4 6 2 7 0 1 2 Full 4 8 5 5 6 3 1 2 Full 4 5 5 2 5 9 Full 3 5 4 2 4 8 Full 3 4 3 9 4 4 17 ft3 Empty 6 5 7 5 8 7 17 ft3 Empty 6 2 7 1 8 1 1 2 Full 5 5 6 3 7 3 1 2 Full 5 3...

Page 13: ...t Point and simultaneously press and hold the Set Point Display button The temperature display changes to read the existing set point value 2 Turn the set point screw clockwise for a colder setting an...

Page 14: ...inet 2 Turn off the freezer 3 Open the outer door and all inner doors 4 Let the freezer stand with doors open for at least 24 hours This allows both the interior and foamed refrigerant system to warm...

Page 15: ...a virtually unlimited standby life unless they develop a leak These containers hold liquid at room temperature and high pressure Several bottles can be connected together increasing the effective len...

Page 16: ...oxes for all chest models except 3 1 and 4 5 cu ft 6230 Single split rack holds 4 3 in boxes for all chest models except 3 1 and 4 5 cu ft 6231 Basket Rail Kit for 6 8 cu ft chest freezers 6217 Basket...

Page 17: ...rogen cylinder Description Catalog No Temperature recorder 6 in circular chart seven day drive panel mounted for 40 C to 95 C Not available for Harris Classic chest models which use 6383 6 Factory ins...

Page 18: ...rdering 5612 Deluxe Electronic Remote Alarm System User programmable to sound alarm in the event of temperature rise or power failure Can dial up to four telephone numbers to advise of alarm condition...

Page 19: ...die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi...

Page 20: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 21: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 38257H01 Rev J...

Reviews: