2

4 Installation

Do not exceed the electrical and temperature ratings printed on 
the dataplate located on the lower left side of the unit.

CAUTION!

Improper operation of the equipment 

could result in dangerous conditions. To preclude 
hazard and minimize risk, follow all instructions and 
operate within the design limits noted on the dataplate.

4.1 Location

Install the unit in a level area free from vibration with a minimum 

of six inches of space on the sides, rear, and top. Refer to 
Section 4.4 for further instructions on leveling cabinets. Allow 
enough clearance so that door or lid can swing open at least 

90 degrees.

Do not position the equipment in direct sunlight or near heating 
diffusers, radiators, or other sources of heat. The ambient 

temperature range at the location must be 59 to 90

°

(15 to 32

°

C).

CAUTION!

To allow for proper air flow, a minimum 

of six inches of clearance space is required behind the 
freezer. 

4.2 Wiring

CAUTION!

Connect the equipment to the correct 

power source. Incorrect voltage can result in severe 
damage to the equipment.

DANGER!

For personal safety and trouble-free 

operation, this unit must be properly grounded before 
it is used. Failure to ground the equipment may cause 
personal injury or damage to the equipment. Always 
conform to the National Electrical Code and local 
codes. Do not connect the unit to overloaded power 
lines.

Your freezer is equipped with one of five NEMA style plugs 

(refer to Figure 1). These plugs meet UL requirements.

Figure 1.  NEMA Style Plugs

Always connect the freezer to a dedicated (separate) circuit. Each 
freezer is equipped with a service cord and plug designed to 
connect it to a power outlet which delivers the correct voltage. 

Supply voltage must be 10%, -5% of the freezer rated 
voltage.

CAUTION!

Never cut the grounding prong from the 

service cord plug. If the prong is removed, the 
warranty is invalidated.

4.3

Voltage Safeguard

Freezers are equipped with a standard built-in Voltage Safeguard 
circuit to detect and adjust low line voltage. 

The Voltage Safeguard helps protect electrical equipment from 
damage by boosting low voltage automatically. On Harris DLT 
and SLT models a panel light flashes to indicate when deficient 

line voltage is being corrected (see Figure 3 on page 4).

4.4 Leveling

The unit must be level both front to back and side to side. 

To level chest models, leave the casters in place and shim the low 

wheel(s) with strips of sheet metal cut at least 1/2 in. wider than 
the caster.

Upright models have adjustable leveling feet which are located 

adjacent to the front casters.

4.5 Door Operation (Upright Models)

Upright freezer models are equipped with a special handle cam 
assembly which provides ease in opening and closing operations.

4.5.1 Locking the Door

1. Close the freezer door and place the handle in a vertical 

position.

2. Insert the key horizontally into the lock.
3. Push the key and lock assembly into the handle 

approximately 0.25 inch and turn the key clockwise 90 
degrees.

4. Continue to depress the key and turn it counterclockwise 

back to the starting position. The lock assembly should 
remain depressed inside the handle.

5. Remove the key.

4.5.2 Unlocking the Door

1. Insert the key horizontally into the lock.

2. Turn the key 90 degrees clockwise. The lock assembly 

automatically springs out until its face is outside the outer 
handle casting).

3. Turn the key back to horizontal (counterclockwise 

90 degrees) and remove the key.

4.5.3 Opening the Door

1. The door must be unlocked (the face of the lock is flush with 

the outer handle casting and not indented into the casting).

2. Grab the upper handle on the front of the freezer and pull it 

toward you while rotating it 90 degrees until the handle rests 
on the built-in stop.

3. Use the vertical door handle mounted on the left front corner 

of the door to pull the freezer door fully open.

15A-125V

NEMA 5-15P

20A-125V

NEMA 5-20P

15A-250V

NEMA 6-15P

20A-250V

NEMA 6-20P

30A-250V

NEMA 6-30P

Summary of Contents for DLT

Page 1: ...Installation and Operation Manual Thermo Scientific Harris DLT SLT and ELT Ultra Low Temperature Freezers...

Page 2: ...d trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights reserved Suva is a regist...

Page 3: ...ocking the Door 2 4 5 2 Unlocking the Door 2 4 5 3 Opening the Door 2 4 6 Compressor Mounts 3 4 7 Pressure Equalization Port 3 5 Operation Harris DLT and SLT Series Freezers 4 5 1 Control Panel Featur...

Page 4: ...arm 9 8 Maintenance and Troubleshooting 10 8 1 Condenser Maintenance 10 8 1 1 Cleaning the Condenser 10 8 1 2 Cleaning the Condenser Filter 10 8 2 Gasket Maintenance 10 8 3 Defrosting the Freezer 10 8...

Page 5: ...the compressors to operate for a prolonged time period Attempting to utilize this freezer improperly may jeopardize user product safety or cause undue stress or damage to the refrigeration compressors...

Page 6: ...the prong is removed the warranty is invalidated 4 3 Voltage Safeguard Freezers are equipped with a standard built in Voltage Safeguard circuit to detect and adjust low line voltage The Voltage Safeg...

Page 7: ...qualization Port When an upright ultra low temperature freezer door is opened room temperature air rushes into the storage compartment When the door is closed the fixed volume of air is cooled rapidly...

Page 8: ...rease temperature values 18 Increment Pad Use this pad to increase temperature values The following indicators are available only when the freezer is equipped with the optional backup system refer to...

Page 9: ...the Warm Alarm Set Point 4 Release both pads when the digital temperature display window shows the correct set point value When the cabinet temperature drops below the Warm Alarm setting turn the key...

Page 10: ...serve the following precautions when using liquid CO2 and LN2 backup systems WARNING If a CO2 or LN2 cylinder falls and a valve is knocked off the cylinder becomes a deadly and completely unguided mis...

Page 11: ...wer WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 4 System Status Lights In addition to the backup control system the built in backup system includ...

Page 12: ...ty 5 6 6 5 7 5 14 ft3 Empty 5 4 6 2 7 0 1 2 Full 4 8 5 5 6 3 1 2 Full 4 5 5 2 5 9 Full 3 5 4 2 4 8 Full 3 4 3 9 4 4 17 ft3 Empty 6 5 7 5 8 7 17 ft3 Empty 6 2 7 1 8 1 1 2 Full 5 5 6 3 7 3 1 2 Full 5 3...

Page 13: ...t Point and simultaneously press and hold the Set Point Display button The temperature display changes to read the existing set point value 2 Turn the set point screw clockwise for a colder setting an...

Page 14: ...inet 2 Turn off the freezer 3 Open the outer door and all inner doors 4 Let the freezer stand with doors open for at least 24 hours This allows both the interior and foamed refrigerant system to warm...

Page 15: ...a virtually unlimited standby life unless they develop a leak These containers hold liquid at room temperature and high pressure Several bottles can be connected together increasing the effective len...

Page 16: ...oxes for all chest models except 3 1 and 4 5 cu ft 6230 Single split rack holds 4 3 in boxes for all chest models except 3 1 and 4 5 cu ft 6231 Basket Rail Kit for 6 8 cu ft chest freezers 6217 Basket...

Page 17: ...rogen cylinder Description Catalog No Temperature recorder 6 in circular chart seven day drive panel mounted for 40 C to 95 C Not available for Harris Classic chest models which use 6383 6 Factory ins...

Page 18: ...rdering 5612 Deluxe Electronic Remote Alarm System User programmable to sound alarm in the event of temperature rise or power failure Can dial up to four telephone numbers to advise of alarm condition...

Page 19: ...die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi...

Page 20: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Page 21: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 38257H01 Rev J...

Reviews: