3 - Instrument Setup
27
Use the fine adjusting knobs to orient the specimen on the x- and y-axes.
Note
The fine adjustment knobs will click into place when the center position has been reached.
Tighten the specimen head clamping lever again.
For further specimen orientation, the specimen chuck can also be rotated by 360° as described below:
Loosen the specimen head clamping lever slightly.
Rotate the chuck as required.
Move the specimen head clamping lever back to its original position.
After the specimen is adjusted, set the cutting window limits - see Setting the Cutting Window.
Use the motorized coarse feed to position the blade close to the specimen.