2 - Introduction
7
Cooling Specifications
Specimen Temperature Control
-50°C to +10°C (-58°F to +50°F) at +20°C room Temp. (decrease
0,7°C/1°C room temp. increase)
Blade Temperature Control
-35°C to -5°C (-31°F to +23°F) at 20°C room Temp (decrease
0,5°C 71°C room temp. increase)
Cryobar Cooling
Max. -60°C (-76°F) at -35°C blade temperature
Defrosting Options
Programmed / Immediate
Microtome and Sectioning Specifications
Section Thickness Range
0.5 – 500 μm
Fine Section Thickness Range
0.5 - 100 μm
Fine Section Thickness Resolution
0.5 – 2 μm in 0.5 μm steps
2 μm – 10 μm in 1 μm steps
10 μm – 20 μm in 2 μm steps
20 μm – 50 μm in 5 μm steps
50 μm – 100 μm in 10 μm steps
Trimming Thickness Range
5 - 500 μm
Trimming Thickness Resolution
5 – 10 μm in 5 μm steps
10 μm – 100 μm in 10 μm steps
100 μm – 200 μm in 20 μm steps
200 μm – 500 μm in 50 μm steps
Specimen Retraction (Return Travel)
20 μm
Vertical Specimen Stroke
64 mm
Horizontal Specimen Movement
48 mm
Specimen Approach
Manual (variable speed by joystick control)
Chuck Size
30, 40 mm (additional special sizes are available)
Specimen Orientation
x - and y - axes universal 7° with zero positioning; z - axis up to
360°
Coarse Feed
Motorized
Cutting Drive
Motorized, electronically controlled and manual via handwheel
Cutting Window
Manually adjusted to specimen size
Operating Modes
Interval, single, multi- and continuous-stroke
Cutting Speed
0 - 256 mm/s (0-120 1/min)
Clearance Angle Adjustment
Disposable blade carrier 8- 16°
Vacutome Filter
approximately 500 ml
Summary of Contents for CryoStar NX50
Page 1: ...Thermo Scientific CryoStar NX70 NX50 Operator Guide 387928 ...
Page 4: ......
Page 6: ......
Page 10: ......
Page 14: ...4 Product Return Safety Declaration ...
Page 120: ...110 Appendices ...
Page 125: ...Index 115 W Waste Tray 16 ...
Page 126: ...116 ...