94
Cold D
Starting the Disinfection (Cold D) Function
The Cold D function allows the user to disinfect the cryo chamber with Sanosil®. Disinfection is either on-
demand or pre-programmed.
The Cold D cycle takes 50 minutes to complete.
T
HE WINDOW OF THE CRYO CHAMBER MUST ALWAYS BE CLOSED BEFORE
STARTING AND DURING A
C
OLD
D
CYCLE
.
Note:
If the window is not closed, the warning “Window not closed” will be displayed.
WARNING:
B
EFORE STARTING THE
C
OLD
D
FUNCTION
,
REMOVE ALL SPECIMENS FROM THE
CRYO CHAMBER
.
Press the “Disinfect” button on the Main screen.
The Disinfection screen will appear.
Press "OK" to start the disinfection immediately.
Press the "Back" button to return to the Main screen without starting the disinfection.