Current Output 2 sub-function
In this sub-function you set the current output range of the transmitter for
temperature values.
ENT
ENT
ENT
ENT
ENT
ENT
ENT
o
c
o
c
o
c
o
c
2
2
3
3
4
4
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
1
cur.2
SETUP
HOLD
1. Select the “CUR.2” sub-function, then
press the ENT key
.
2.
Selecting the output type: press the
▲
or
▼
key
to select the desired
output type: 0-20 mA or 4-20 mA.
Press the ENT key
to confirm your
selection.
3.
Setting the temperature value at which the transmitter output will be
4 mA (or 0 mA): press the
▲
or
▼
key
to set the temperature value
(setting range:
-9.9 to 115°C or 15 to 237 °F) to be equivalent
to 4 mA (or 0 mA).
Press
the ENT key
to confirm your setting.
4.
Setting the temperature value at which the transmitter output will be
20 mA:
press the
▲
or
▼
key
to set the temperature value (setting
range: -0.1 to 125°C or 35 to 257 °F) to be equivalent
to 20 mA.
Press the
ENT key
to confirm your setting.
Continue with Setup mode procedures, or return to Measurement mode by
pressing the
▲
and
▼
keys simultaneously (escape).