8. Resetting the controller settings to factory defaults: The display shows “NO
DEF”.
Press the
▲
or
▼
key
to select:
– “NO DEF” = keeps old values active, when confirmed with ENT.
– “FCT DEF” = resets all settings to factory defaults, when confirmed with
ENT.
– “CAL DEF” = resets calibration settings to factory defaults, when confirmed
with ENT.
Press the ENT key
. The selected function will be executed and the controller
returns automatically to the Measurement mode.
Continue with Setup mode procedures, or return to Measurement mode by
pressing the
▲
and
▼
keys
simultaneously (escape).
6.9
Calibration (CAL) sub-function
The calibration procedure in Setup mode is identical to the procedure in the
Calibration mode (see section 5). The only difference is that the transmitter
remains in Setup mode (instead of Measurement mode) after calibration is
completed.