6.2 Electrode Offset Settings (OFS)
The electrode offset settings allow you to zero the transmitter without a
sensor. The use of a simulator is recommended. For real zero calibration,
use the calibration procedure as described in section 5. If no simulator is
used, a 22 k
Ω
NTC must be connected to pins 24 and 25 of the terminal
connector on the backside of the unit (see section 3.2).
SETUP
HOLD
OFS
SETUP
HOLD
OFS
SETUP
HOLD
SETUP
HOLD
BAR
PRES
PRES
PRES
SETUP
HOLD
H9
SETUP
HOLD
PSI
1.013
BAR
SETUP
HOLD
NO
OFS
OFS
SETUP
HOLD
YES
SETUP
HOLD
0.0
0.0
ppt
%
SAL
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
This screen will not be
visible if DO Unit of
Measurement is set to %
(In Configuration settings)
1
3
2
2
5
6
6
6
7
SETUP
HOLD
ADJ
0.0
%
ENT
4
1. From the setup mode select the “OFS” screen and then
press ENT key
.
2.
Selecting to modify offset value:
By default,
controller
shows “NO OFS”.
Press
▲
or
▼
key
to select “NO” or “YES” for offset adjustment.
−
Select “NO OFS” to skip offset adjustment and move to the next step.
−
Select “YES OFS” to perform offset adjustment.
Press ENT key
to confirm your selection.
−
If “YES OFS” was selected:
3.
Selecting offset value
: The display shows the offset from zero current of
the saturation or concentration. If this is within the range of -0.2 nA to 1.1 nA,
then the correction can be accepted.
Press ENT key
to confirm.
4. The display shows the offset.
Press ENT key
to confirm.
5.
Setting salinity of the solution (SAL):
This display only appears if the
measurement unit is set to “mg/l” or “ppm”.
Press
▲
or
▼
key
to set the
salinity of the solution.
Press ENT key
to confirm.
6.
Setting the unit for pressure (PRES): Press the
▲
or
▼
key
to set unit
(“BAR”, “mmHG”, “PSI”) of the barometric pressure.
Press the ENT key
to
confirm.
7.
Setting barometric pressure: Press
▲
or
▼
key
to set the barometric
pressure according to the selected unit in step 6.
Press ENT key
to confirm.
Continue with rest of the Setup procedures by pressing
▲
or
▼
key or return to
the Measurement mode by pressing
▲
and
▼
keys (escape) simultaneously.