Instruction manual
Alpha
DO 2000
seconds (0 to 1999 seconds) you select.
Press ENT key
to confirm your
setting.
Î
NOTE:
You can set a time delay for each relay, which stops the relay from
switching on the moment the set point is exceeded. This controller lets
you set a 0 to 1999 seconds time delay before your relay activates.
6.
Setting the off-delay time lag: press
▲
or
▼
key
to enter the off-delay time
for set point 1. The controller will delay deactivation of the relay for the number
of seconds (0 to 1999 seconds) you select.
Press ENT key
to confirm your
setting.
Î
NOTE:
You can set a time delay for each relay, which stops the relay from
switching off the moment the value reaches the set point and
hysteresis. This controller lets you set a 0 to 1999 seconds time delay
before your relay deactivates.
The controller goes back to “SP1” screen.
Continue with rest of the Setup procedures by pressing
▲
or
▼
key or return
to the Measurement mode by pressing
▲
and
▼
keys (escape)
simultaneously.
23