Safety Tips

1. Use independent power supply.
2. Check if the local power supply 
     voltage is suitable for use.
3. Do not drag the power supply cable 
     when unplugging.
4. Do not use non-specified power cable 
     or damage cable.
5. Service should only be performed by a 
     qualified professional.
6. The power supply must be unplugged 
     under the following situations:
     (1). When the unit is moved
     (2). When the electrical cabinet or the 
            moving component is opened
     (3). When the equipment is 
            malfunctioning
     (4). When the equipment is not in use

Maintenance

Instrument case and tray surface can be 
cleaned with cloth with mild detergent 
and water. Surface of the dimpled rubber 
mat can be cleaned with mild detergent 
or water and then wiped with clean cloth.

Warning:

 Avoid dripping detergent or 

water into the inside of the instrument 
during cleaning.

Clean Spill

If accidental spillage of liquids caused by 
mishandling or contained breakage 
occurs on the surface of the instrument, 
please shut down the instrument and 
clean up the liquid immediately.
If the liquid has already spilled into the 
unit, cut off the power supply first and 
immediately clean up the liquid at the 
surface of the instrument. Place the 
instrument in a ventilated and dry 
environment for 24 hours before reuse. If 
the instrument is not functioning after 
drying for 24 hours, please contact the 
manufacturer.

Warning:

 Disassembling/Assembling 

without a qualified professional's 
guidance may cause malfunctioning of 
the instrument.

11        Section 5  |  Safety Tips and Maintenance

Digital Rocker     12

Section 5 Safety Tips and Maintenance

Section 6 Troubleshooting

Error

Cannot start

instrument, LED

display window

off

No shaking

of the tray

Loud noise

Cause

Power disconnected

Power switch off

Power adaptor failure

Over-weighted or
unbalanced load

Electrical malfunction 

Mechanical malfunction

Tray loose

Instrument vibrate

Housing loose

Solution

Connect the power

Switch on power

Replace power adaptor

Adjust the weight and position
of load, decrease rotation speed

Contact Thermo Scientific

Contact Thermo Scientific

Fasten screws

Place instrument at a flat and
firm suface

Fasten housing screws

Other

Keep record for maintenance

Please refer to the following table to 
troubleshoot if any malfunction occurs.

If the problem still exists, contact your 
local sales representative.

Shut down and restart, useless,
please contact Thermo Scientific

Err1

Operational obstruction

Summary of Contents for 88882001

Page 1: ...fisher com In Canada For customer service call 1 800 234 7437 To fax an order use 1 800 463 2996 To order online thermofisher ca For Research Use Only Not for use in diagnostic procedures 2019 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified Operating Manual Revision A 09 03 2019 ...

Page 2: ... Section 3 Overview Specifications Environmental Conditions Safety Instructions Loads Angle and Speed Section 4 Operation Control Panel Installation Settings Section 5 Safety Tips and Maintenance Section 6 Troubleshooting Section 7 Optional Accessories Section 8 Warranty Information Contents 1 2 3 3 4 4 5 5 5 6 7 8 8 9 9 12 13 14 15 ...

Page 3: ...looding can happen Do not assemble repair modify on your own The product may not work well and electric shock is possible with changes in the efficiency of the product Also this will void the warranty Indicates a hazardous situation which if not avoided may result in minor or moderate injury Do not put heavy things on the power cord Do not put the machine on the cord It may take off the wire coati...

Page 4: ...heck carefully to see if there were damages incurred during transit After unpacking check that all unit parts and accessories are as listed below Contact us or the agent from which you purchased the unit if any components were omitted 3 Section 2 Inspection and Installation Digital Rocker 4 General Power Adaptor US Plug CN Plug EU Plug UK Plug Catalog Number 88882001 1 10 N A N A N A Digital Rocke...

Page 5: ...leaning DO NOT operate the unit if it shows signs of electrical or mechanical damage WARNING DO NOT use the in a hazardous atmosphere or with hazardous materials for which the unit was not designed Also the user should be aware that the protection provided by the equipment may be impaired if accessories used are no provided or recommended by the manufacturer or are used in a manner not specified b...

Page 6: ... Time Tilt display window The window displays cumulative time in continuous mode or remaining time in timer mode The range of time displayed is 0s to 9999 minutes Time indicator The light is on when the instrument is running in Timer Mode Tilt indicator The light is on when the instrument is in Tilt Adjustment Mode Mode button Switch modes between time and Tilt settings Time Tilt setting buttons U...

Page 7: ...strument according to the storage guide Alarm System Err1 If running speed does not match the setting speed the instrument will have a three time reboot and after the three reboots failed there will be a buzzer and then Err1 will be displayed in the Speed display window End of timer The instrument buzzing alarm the End is displayed in the Speed window When the instrument alarms press any key the i...

Page 8: ... immediately clean up the liquid at the surface of the instrument Place the instrument in a ventilated and dry environment for 24 hours before reuse If the instrument is not functioning after drying for 24 hours please contact the manufacturer Warning Disassembling Assembling without a qualified professional s guidance may cause malfunctioning of the instrument 11 Section 5 Safety Tips and Mainten...

Page 9: ...epairs Expendable items glass filters and gaskets are excluded from this warranty Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non ...

Reviews: