4
Rotor Loading
Maximum Loading
Thermo Scientific
75003623 CLINIConic 30
x
15
4-3
Maximum Loading
The rotor can run at high speeds. The rotor design has sufficient reserve stability even when spinning at
top speed.
The safety system of the centrifuge requires that you do not overload the rotor.
There are two options available for centrifuging samples whose weight, including adaptor, exceeds the
maximum permissible load:
•
Reduce the fill level.
•
Reduce the speed.
Use the table or formula below.
Recalculate the maximum permissible speed according to the following formula and select the
calculated maximum speed for the centrifuge:
n
adm
= admissible speed
n
max
= maximum speed
Actual Load (g)
RPM
max
30
5650
50
4376
70
3699
90
3262
110
2951
130
2714
150
2527
170
2373
n
adm
n
max
Maximum permissible load
Effective load
----------------------------------------------------------------------
=