4
Rotor Loading
Before a Run
4-2
75003623 CLINIConic 30
x
15
Thermo Scientific
Before a Run
1.
Please read and observe the safety instructions contained in these operating instructions and in
the instructions for use.
2.
Check the rotor and all accessory parts for damages such as cracks, scratches or traces of
corrosion.
3.
Check the rotor chamber, the motor shaft and the AutoLock
™
.
4.
Check the rotor’s suitability using the
“Chemical Compatibility Chart”
5.
Make sure the tubes or bottles do not touch the lid.
Proper Loading
To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times.
All places must contain stainless steel tube holders. Running the rotor with missing stainless steel
tube holders is forbidden. Start loading the rotor on the outer ring. If you only centrifuge one sample
place a test tube with water in the opposite position to balance the load.
Improper Loading