Figure 3-5.
Time Dispense Mode Operation
1. Getting Started: From the Main Menu, use the enter key to select Time
Dispense Mode to enter the Time Dispense Mode Operation screen.
2. Calibrating Tubing: Before operating the pump, insert desired tubing
into the Pump Head. For more information, see “
Tubing Calibration
”.
3. Choosing Settings: Select desired flow unit, tube size, flow rate, pump
direction, etc. For more information see “
SETUP Menu
.”
4. Preparing Tubing: Insert tube inlet into supply fluid. Next, insert tube
outlet into desired container.
5. Selecting Flow Rate: Use the directional pad and ENTER key to select
the Setup icon. Use the UP/DOWN arrows on the directional pad to
select Flow Rate. In the Flow Rate selection screen, press the ENTER
key and then use the UP/DOWN arrows on the directional pad to
select a desired flow rate. For faster entry, use the LEFT/RIGHT
arrows on the directional pad to move between digits and the
UP/DOWN arrows to increase or decrease the value, respectively.
Press ENTER one more time to validate the selected flow rate. Use the
directional pad to select EXIT to return to the Time Dispense Mode
Setup Screen. Select EXIT again to return to Time Dispense Mode.
6. Setting ON Time: To set the ON Time, use the directional pad and
ENTER key to select the ON field (see position D, Figure 3-5). Doing
so will highlight the timer in the center of the screen (see position I,
Figure 3-5). Pressing ENTER again, allows the timer to be set using
the UP/DOWN arrows. Switch between digits using the
LEFT/RIGHT arrows. Having selected an optimal ON Time, press
ENTER again to validate. The drive will now run for the time
appearing in the center of the screen.
NOTE:
When entering the Time Dispense Mode Screen the ON or OFF
Field will be highlighted (see position D or E, Figure 3-5). The highlighted
field indicates whether the time displayed is for the ON time or OFF
Time.
Time Dispense Mode
Operation
TIME DISP. MODE
00:00:00
OFF
ON
1000/2000
I
D
E
Section 3
Operation
Thermo Scientific
3-10
FH100D/FH100DX Dispensing Pump Operating Manual