Servicing
DMC Optical Bench Removal
Thermo Scientific
43iQ Instruction Manual
7-31
Use the following to remove the optical bench assembly.
Equipment required:
Hex nut drive, 3/8 and 5/32 and hex drive, 5/32
1.
Unplug all electrical as shown (
Figure 7–35.
Removing the Optical Bench Assembly
2.
Using 3/8 hex nut drive, unfasten nut.
3.
Using 5/32 hex drive, unfasten two #10-32 socket cap head screws.
4.
Pull optical bench assembly upwards (
Optical Bench