If the alarm temperature probe fails these tests as shown in Chart 3-3, the
probe is either defective or has become unplugged from the circuit board
(Figure 2-4). Contact the Technical Services department at 1-888-217-
1790, if necessary.
Pressing the center button below the
display screen will silence the audible
alarm. Refer to Figure 3-8. However,
the alarm will ring-back in 30 minutes
if the alarm condition still exists.
Test the performance alarm battery by
turning off the unit or pulling the plug
from the electrical outlet. The audible
alarm should sound and the Power
Failure indicator should light. Allow
the alarm to continue for 5 to 10
seconds before returning power to the
unit.
If the alarm does not activate, replace
the battery with stock number
400116. Refer to Figure 3-9 and to the
installation instructions in Section 2.
Perform the alarm test at intervals recommended by the laboratory or the
bolld bank industry.
7033881 Blood Bank 3-7
Thermo Scientific
Section 3
Operation
If the probe tests fail
Silencing the Audible
Alarm
Figure 3-8.
Testing Performance
Alarm Battery
Figure 3-9.
Summary of Contents for 3881
Page 9: ......
Page 19: ......
Page 23: ...3 4 7033881 Blood Bank Thermo Scientific Section 3 Operation ...
Page 27: ...3 8 7033881 Blood Bank Thermo Scientific Section 3 Operation ...
Page 28: ...7033881 Blood Bank 3 9 Thermo Scientific Section 3 Operation ...
Page 29: ......
Page 35: ......
Page 37: ......
Page 39: ......
Page 40: ...7033881 Blood Bank 8 1 Thermo Scientific Section 8 Refrigeration Schematics ...
Page 41: ...8 2 7033881 Blood Bank Thermo Scientific Section 8 Refrigeration Schematics ...
Page 42: ...7033881 Blood Bank 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Page 43: ...9 2 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 44: ...7033881 Blood Bank 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Page 45: ......
Page 46: ...7033881 Blood Bank 9 4 Thermo Scientific Section 9 Electrical Schematics ...
Page 47: ...9 5 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 48: ...7033881 Blood Bank 9 6 Thermo Scientific Section 9 Electrical Schematics ...
Page 49: ......
Page 50: ...7033881 Blood Bank 9 7 Thermo Scientific Section 9 Electrical Schematics ...
Page 51: ...9 8 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 52: ...7033881 Blood Bank 9 9 Thermo Scientific Section 9 Electrical Schematics ...
Page 53: ......
Page 54: ...7033881 Blood Bank 9 10 Thermo Scientific Section 9 Electrical Schematics ...
Page 55: ...9 11 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 56: ...7033881 Blood Bank 9 12 Thermo Scientific Section 9 Electrical Schematics ...
Page 57: ......