7033881 Blood Bank 5-1
Thermo Scientific
Section 5
Troubleshooting and Service
The following is a guide for troubleshooting some of the common
problems that may be encountered over the operating life of the
refrigerators. This section is provided as a support to qualified electrical
and refrigeration service.
Symptom
Possible Cause/Possible Fix
Refrigerator appears to be completely inoper-
ative; audible alarm is activated and there is
no green light.
Power line cord is disconnected.
Power source circuit breaker is open or
source fuse is blown.
Power switch is not turned on (3883, 3887).
High temperature indicator and audible alarm
are activated after the door is opened and
are deactivated shortly after the door is
closed.
Alarm system probe is defective.
Note
This situation may be normal if the
alarm probe is not immersed in liquid.
Temp control needs adjusted.
Alarm point may need adjusted.
Low temperature indicator and audible alarm
are activated intermittently or continuously.
Chamber temperature is unstable and may
occasionally be below 1.5°C (+/-0.5°C).
Refrigeration system temperature controller
may need adjusting.
Alarm point may need adjusted.
Low temperature indicator and audible alarm
are activated intermittently or continuously;
chamber temperature is relatively stable at
4°C.
Probe or control board may be defective.
Alarm points may need adjustment.
Possible loose probe wires.
Summary of Contents for 3881
Page 9: ......
Page 19: ......
Page 23: ...3 4 7033881 Blood Bank Thermo Scientific Section 3 Operation ...
Page 27: ...3 8 7033881 Blood Bank Thermo Scientific Section 3 Operation ...
Page 28: ...7033881 Blood Bank 3 9 Thermo Scientific Section 3 Operation ...
Page 29: ......
Page 35: ......
Page 37: ......
Page 39: ......
Page 40: ...7033881 Blood Bank 8 1 Thermo Scientific Section 8 Refrigeration Schematics ...
Page 41: ...8 2 7033881 Blood Bank Thermo Scientific Section 8 Refrigeration Schematics ...
Page 42: ...7033881 Blood Bank 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Page 43: ...9 2 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 44: ...7033881 Blood Bank 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Page 45: ......
Page 46: ...7033881 Blood Bank 9 4 Thermo Scientific Section 9 Electrical Schematics ...
Page 47: ...9 5 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 48: ...7033881 Blood Bank 9 6 Thermo Scientific Section 9 Electrical Schematics ...
Page 49: ......
Page 50: ...7033881 Blood Bank 9 7 Thermo Scientific Section 9 Electrical Schematics ...
Page 51: ...9 8 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 52: ...7033881 Blood Bank 9 9 Thermo Scientific Section 9 Electrical Schematics ...
Page 53: ......
Page 54: ...7033881 Blood Bank 9 10 Thermo Scientific Section 9 Electrical Schematics ...
Page 55: ...9 11 7033881 Blood Bank Thermo Scientific Section 9 Electrical Schematics ...
Page 56: ...7033881 Blood Bank 9 12 Thermo Scientific Section 9 Electrical Schematics ...
Page 57: ......