Model 916/917_____________________________________________________________________________________________________Preventive Maintenance
3-3
Preventive Maintenance for 900 Series Freezers
Refer to Manual Section
Action
Monthly
Yearly
Every
2 Years
--
Verify ambient temperature, <90°F
;
--
* Adjust door handle for firm latching, as needed
;
Figure 1-4 probe location 4.5
Check and clean probe and cover, gaskets, hinges and doors of ice and snow.
;
More frequent cleaning may be
required, depending on use and
environmental conditions.
4.2
Check air filter. Clean or replace as needed
;
1.7
Check alarm back-up battery.
;
** Replace
--
Check condenser fan motor for unusual motor noise or vibration.
;
--
* Verify and document calibration, at the minimum, annually.
;
4.3
* Clean condenser compartment and wipe off condenser
;
* Qualified service technicians only
** Dispose of properly, according to all state and federal regulations.